5 Others [
①
Select User Cal]
(1) Select “
Select User Cal
” on the extension function and press
the
key.
(2) Use the
▲
or
▼
key to select from “
User0
” through “
User7
”
or “
S & D
” and press the key.
* When the user calibration has not been completed, the calibra-
tion cannot be selected. Refer to “
②
Make User Cal” on page 19
for how to set the calibrations.
* “D.MODE” and “S.MODE” are displayed at the same time for the
S & D mode.
(3) Press the key to return to the measurement screen on the
calibration selected.
• Description on D mode/S mode
The D mode is displayed not with the moisture content but with
the value of 0 to 1999, which correlates to the high-frequency
capacitance signal. The relationship between moisture con-
tents and the D mode is shown as the graph at right. Measured
values equivalent to the D.MODE of HI-520 will be provided.
Moisture content (%)
D mode