28
8-3 Selection of UNIT (measurement standard and minimum display digit)
● Selecting the measurement standard
This is used to select and specify the measurement standards to be used when performing measurements. There
are 3 types of measurement standards; measuring by changes in wet content, measuring by changes in dry
content, and measuring by changes in solid content, and therefore, the standard to be used should be selected in
accordance with the type of material to be measured.
Type
Display during specification
of setting
Formula
Description
Wet Base
W – D
W
×100 (%)
Percentage of evaporated
moisture weight with respect to
the weight before drying.
Dry Base
W – D
D
×100 (%)
Percentage of evaporated
moisture weight with respect to
the weight after drying.
Solid
D
W ×100 (%)
Percentage of residual weight
after drying with respect to the
weight before drying.
[Notation used in formula]
W : Wet weight before measurement, D: Dry weight after measurement
(While a measurement is being performed, the weight at each point in time is used as the dry weight in calculating
measurements.)
● Selecting the minimum display digit
Select 0.1% or 0.01% for the minimum display digit, and configure it.
* Note that the minimum display digit has no effect when performing comparative measurements.
The UNIT setting is common for all measuring setting storage areas.
● Setting procedure
1
With the display showing the weight in grams, press the
[
Menu
] key.
2
[Menu] lights up, and the number of the current measuring
setting storage area blinks.
The UNIT setting is common to all measuring storage
areas, and therefore, selection is not required.
Just press the [
Enter
] key.
3
①
②
Move to the menu item selection.
Every time the [
Select
] key is pressed, the menu items
alternately blinks in the order of “MODE” → “TEMP” →
“UNIT” → “OUTPUT” → “BIAS” → “CAL” → “MODE” → …
①
Press the [
Select
] key.
②
Press the [
Enter
] key after “UNIT” blinks.
* For UNIT, the measuring setting storage areas
1
to
5
are specified in common.