background image

21

● Setting procedure

1

With the display showing the weight in grams, press the 

[

Menu

] key.

2

[Menu] lights up, and one of 

1

 to 

5

 blinks.

Select an area number to which the measuring settings 

are stored from 

1

 to 

5

 with use of the [▲] or [▼] key.

3

After the desired number blinks, press the [

Enter

] key.

“3” is selected in the example on the left.

4

The "3" changes from blinking to lit.

Now, the measuring setting storage number is set to 3.

At the same time, "MODE" in the menu blinks, and the 

menu  setting  can  now  be  performed  (see  "8.  Menu 

Settings" on page 22).

Pressing the [

Menu

] key returns the screen to the weight 

display screen.

5

Summary of Contents for FD-660

Page 1: ...Infrared Moisture Analyzer FD 660 Operating Manual Thank you for purchasing this product Please read the operating manual carefully and use this product properly...

Page 2: ......

Page 3: ...rement of liquid sample 19 Measurement of large particulate sample 19 Removal of soil sample gas and others 19 Deodorizing windshield case 19 7 Specifying Measuring Settings 20 8 Menu Settings 22 8 1...

Page 4: ...e if flammable materials are placed nearby This product should be used only with the applicable power supply voltage Using this product with incorrect voltage may cause the product to be overheated wh...

Page 5: ...an odd smell or any other sign of abnormal functioning remove the power plug from its socket and take whatever other steps would be appropriate to deal with the problem Notes on using the control key...

Page 6: ...rvice life 2 types of measuring modes This product provides 2 types of measuring modes and therefore can perform measurement under appropriate drying settings in accordance with the drying characteris...

Page 7: ...ta logger software FDL 02 option Storage of measurement settings 5 types Temperature humidity op erating range 5 40 C 85 RH or less no condensation Heat source Organic carbon heater 280 W x 2 Temperat...

Page 8: ...8 4 Part Names 4 1 Names of Main Unit Parts Heater lid Heater Sample dish Wind shield Display and control panel Power switch Power conversion switch AC inlet RS 232C output terminal...

Page 9: ...nd A 4 Spoon A 5 Spare fuse 8A 2 A 7 Sample dish handler A 2 Aluminium sheets x10 A 6 Power cord A with 3P 2P plug adapter A 8 or Power cord B A 9 Operating Manual A 10 4 2 Accessories The aluminium s...

Page 10: ...ghts up when the screen can be returned to the weight display screen after measurement Start Stop display Lights up when the unit can start measuring Blinks during measurement Lights up when measureme...

Page 11: ...ure 130 C time 5 minutes fixed Replace the sample dish after preheating 3 Menu key Used at the start and end of specifying a measuring mode a drying temperature a measurement standard measurement data...

Page 12: ...is turned on Attaching the wind shield Open the heater lid Place the wind shield so that the mark on the shield points downward When the wind shield is appropriately placed no backlash or rotation ma...

Page 13: ...tlet Connecting the printer option When a printer is used use the printer specific cable that is supplied with the printer Connect the cable to the RS 232C output terminal Refer to the VZ 330 Printer...

Page 14: ...first time ensure that the power supply voltage is correct and press the Enter key How to manually specify the power supply voltage Turn on the power while holding the Menu key down A beep sounds All...

Page 15: ...ng then you should specify the measurement settings see 7 Specifying Measuring Settings on page 20 Zero Point Adjustment Check to make sure that the stability indicator O is being displayed and then p...

Page 16: ...inish during moisture content measurement Never open the heater lid during moisture content measurement Doing so is not only hazardous but also could contribute to making it impossible to obtain accur...

Page 17: ...the Next Measurement Leave the lid to the heater open to allow the tester to cool When performing sequential measurements make sure that the entire unit has cooled before proceeding to the next measur...

Page 18: ...essary Press the Pre Heat key 2 The Start Stop key blinks heating begins It is heated at 130 C 3 Pre heating will finish in 5 minutes The Tare Reset key lights up 4 Press the Tare Reset key PrEH will...

Page 19: ...e aluminium foil sheets are water permeable thus making it possible to obtain wide and even placement and making them effective for use in shortening measuring times and obtaining accurate measurement...

Page 20: ...re and BIAS offset value UNIT measurement standard and minimum display digit and OUTPUT output format are common in all areas The stored settings can be always recalled and therefore storing settings...

Page 21: ...to 5 with use of the or key 3 After the desired number blinks press the Enter key 3 is selected in the example on the left 4 The 3 changes from blinking to lit Now the measuring setting storage numbe...

Page 22: ...rement but the test is completed while there may still be a wide variation in measurements Determine the automatic settings in accordance with your objectives and the type of sample to be measured Set...

Page 23: ...nks Automatic measuring mode is now selected 5 Automatically the monitoring time is displayed The currently selected monitoring time blinks 6 Every time the or key is pressed 0 5 min 1 min 1 5 min 2 m...

Page 24: ...hange the number use the or key see 7 Specifying Measuring Settings on page 20 When the desired measuring settings area number starts flashing press the Enter key 3 Move to the selection of menu item...

Page 25: ...tes is specified in the example on the left Configurable time range 1 min to 120 min Press the Enter key The measuring time is now specified 7 Since the measuring mode is now specified the MODE starts...

Page 26: ...the material s color moisture content type or shape Generally speaking drying is performed more quickly when a higher drying temperature is specified but if the temperature specified is too high it m...

Page 27: ...ified in the example on the left Press the Enter key Configurable temperature range 30 C to 180 C 6 The blinking drying temperature changes from blinking to lit and the drying temperature is specified...

Page 28: ...used in formula W Wet weight before measurement D Dry weight after measurement While a measurement is being performed the weight at each point in time is used as the dry weight in calculating measure...

Page 29: ...The currently selected minimum display digit blinks 6 Every time the Select key is pressed 01 and 001 alternately blink Press the Select key Press the Enter key after the desired minimum dis play digi...

Page 30: ...asurement conditions are not output after second or later measurements Turned on the power again Entered the Menu see 8 Menu Settings on p 22 Printed the signature field see 6 Measurement Procedure 5...

Page 31: ...to the selection of output interval 6 The currently selected output interval blinks Every time the or key is pressed 30 S 1 M 2 M 5 M 10 M FIN 30 S alternately blink 7 Press the or key Press the Enter...

Page 32: ...Insert the connector at the end of the RS 232C cable into the RS 232C port located at the rear of the FD 660 and then tighten the screws located at both sides of the cable connector Follow the same p...

Page 33: ...atic settings setting min delimiter 7 Timed format _Mode_ _Time delimiter _Setting_Temp _ _ XXX 3 byte temperature setting C delimiter _Drying_Time_ _ XXX 3 byte drying time setting min delimiter 8 _B...

Page 34: ...would cause the time required for measurement to become too long or otherwise result in problems a bias offset should be specified instead When using more than one unit there may be times when it is...

Page 35: ...et value 0 2 is specified in the example on the left Press the Enter key 6 The blinking offset value changes from blinking to lit and the offset value is specified BIAS starts blinking To exit from BI...

Page 36: ...ater lid must be closed during calibration to avoid wind effects If you wish to abort a calibration already in progress press the Tare Reset key An Abort message will be displayed and the screen will...

Page 37: ...after 50 000g blinks Blinking turns to lit Press the or key to specify the mass of the weight In the example on the left 50 015 grams is specified Press the Enter key The specified mass of the weight...

Page 38: ...aced so that the center of gravity of the weight is near the center of the dish 6 Close the heater lid and press the Enter key Be sure to close the heater lid 7 g is displayed Now the 50 gram calibrat...

Page 39: ...r lid 10 g is displayed Now 0 point calibration is in progress 11 After a while scale calibration is completed End is displayed If this unit is connected to a printer a calibration record is automatic...

Page 40: ...Maintenance of parts and accessories The spoon sample dish and wind shield can be washed in water with a sponge Remove the sample dish and wind shield from the main unit when washing them Wait until...

Page 41: ...ay deteriorate and its life may be shortened Finally attach the heater lid 1 2 1 Remove the screws on both sides of the heater lid after ensuring that the heater lid and heater are thoroughly cooled 2...

Page 42: ...r Er306 Heater lid open Press the Tare Reset key to cancel the error Er401 Internal weight measuring unit communications error Turn off the power and then turn the power back on again Er501 Invalid we...

Page 43: ...43...

Page 44: ...document may differ from those of the actual products and accessory items although operation and function will not be affected Although the contents of this document were drafted with great care if y...

Reviews: