background image

19

19

Date & Time

The date and time, as well as date and time formats, can be adjusted. The Time Formats available are 12 
hour and 24 hour.  The Date formats available are day/month/year and month/day/year. (See the Date and 
Time Set Up section for instructions.) Press the 

 button to return to the Main Setup Menu.

Language 

Displayed text can be set in one of fi ve languages: English, French, German, Italian or Spanish.  To choose 
a language, use the 

 or 

 buttons to highlight the desired language. Press the 

 button to select the 

language and return to the Main Setup Menu. Otherwise, press the 

 button to return to the Main Setup 

Menu without changing languages.

Restore

Default settings for units of measure, date and time formats, and system settings can be restored. (See the 
Factory Default Settings section for more information.) Press the 

 or 

 button to highlight the desired 

default setting: Metric, Imperial or Defaults.  Press the   or 

 button to reset the factory setting. Press the 

 button to return to the Main Setup Menu.

17

17

that are hidden, simply enter the Measurement screen and turn them back on. The data will be stored in the 
same manner as all other measurements, as specifi ed in the Store Rate.

Graph Scale

These settings control the chart limits of your Kestrel Meter. Depending on the conditions, the lower and 
upper limits of the chart scale may need to be adjusted in order to get the best view of the data. Highlight 
the desired measurement by pressing the 

 or 

 button. Select the highlighted measurement by 

pressing the 

 button. Press the   or 

 button to increase or decrease the value of the limits. Press the 

 

or 

 button to change between the upper and lower limits. Press the 

 button to exit and return to the 

Measurement Selection screen. Press the 

 button to return to the Main Setup Menu.  

Highlight the desired measurement by pressing the 

 or 

 button. Press the   or 

 button to scroll 

through the available units.  Press the 

 button to return to the Main Setup Menu.

User Screens

The three User Screens can be reconfi gured to display the most appropriate information for your 
application. For example, if you need to monitor the wind speed, humidity and barometric pressure, a User 
Screen can display these current measurement values on the same screen for quick reference. Only current 
measurements can be selected for the User Screens — Min/Max/Avg and Charts are not available.

Highlight the desired User Screen by pressing the 

 or 

 button. Press the 

 button to select the 

highlighted User Screen. Press the 

 or 

 buttons to change lines, and the   or 

 button to scroll 

through the available measurements for each highlighted line. Press the 

 button to return to the User 

Screen Setup Menu. Repeat above process for the other User Screens or press the 

 button to return to 

the Main Setup Menu. The User Screens may also be hidden if not needed.

20

Kestrel

®

 4200 Pocket Air Flow

®

 Tracker

APPLICATION EXAMPLES

This section provides examples of applications where a Kestrel Meter might be used, and the appropriate 
memory settings.

HVAC - Environmental Monitoring

Auto Store      

On

Store Rate           

5 min

Overwrite      

On

Man Store     

Off 

These settings will record conditions every fi ve minutes, for a total storage of almost 7 days. You can 
monitor the conditions in a laboratory or manufacturing plant, both day and night, to determine if the 
climate control is working properly.  Or you can examine the eff ect on the environment when employees 
enter and exit the building. 

HVAC/R - System Balancing

Auto Store     

Off 

Store Rate     

Overwrite      

Off 

Man Store     

On 

These settings will require you to press the Manual Store Button in order to store any data at a duct, hood, 
vent, or other system location. The meter will not store any data automatically.  Be sure to record the 
location and date/time of storage for reference when reviewing the data.  After storing the conditions at 
each location, simply review the data and balance the system. 

18

Kestrel

®

 4200 Pocket Air Flow

®

 Tracker

System  

The display Contrast and Auto Shutdown can be reconfi gured as required. The relative humidity and 
pressure sensors can also be recalibrated. Press the 

 or 

 buttons to highlight the appropriate selection, 

and the   or 

 button to adjust or select.

The Contrast can be adjusted for better visibility depending on the ambient lighting conditions. Press the   
or 

 button to increase or decrease the contrast from 0 to 20 (0 is lightest, 20 is darkest).

The display can be set to automatically turn off  in order to conserve the battery life. Auto Shutdown will 
only occur after the preset time has elapsed without any button presses. Press the   or 

 button to scroll 

through the Auto Shutdown options (15 minutes, 60 minutes, Off  ). Battery life will be shortened if the Auto 
Shutdown is turned to Off .   

Baro Cal - 

Recalibration of this sensor is not typically required, and it is not recommended that you recalibrate 

without speaking to an NK technician.

 The pressure sensor can be calibrated if necessary. It is extremely 

important to know the precise altitude and mean sea level barometric pressure at the time of calibrating 
the sensor. First, set the reference altitude on the BARO measurement screen to the known altitude (see 
Pressure Adjustment section for instructions). Then adjust the calibrating setting on the Baro Cal screen to 
the known mean sea level barometric pressure. If recalibration is desired, the unit may also be returned to 
NK for service.   

Humidity Cal - 

Recalibration of this sensor is not typically required, and it is not recommended that you 

recalibrate without speaking to an NK technician.

 The humidity sensor can be calibrated by “teaching” it the 

correct humidity. Some special equipment is required for this calibration, including two hermetically sealed 
containers and saturated salt solutions. NK off ers a calibration kit, and instructions are available on 
www.nkhome.com. If recalibration is desired, the unit may also be returned to NK for service.    

Summary of Contents for 4200

Page 1: ... 25 CUSTOMER SERVICE 27 Congratulations on the purchase of your Kestrel 4200 Pocket Air Flow Tracker The Kestrel 4200 is our newest and most comprehensive HVAC specific instrument It not only measures EVERY major environmental condition easily accurately and right in the palm of your hand but now also automatically calculates Volume Air Flow CFM and Humidity Ratio grains While the Kestrel 4200 is ...

Page 2: ...r 8 Kestrel 4200 Pocket Air Flow Tracker SPECIAL FUNCTIONS User Screens The Kestrel Meter has three User Screens that can be customized to display three current measurements simultaneously These screens can also be turned off See the Main Menu section for setup instructions Max Avg for Air Flow Air Velocity Wind Chill The Max Avg values for Air Flow Velocity and Wind Chill are measured independent...

Page 3: ...t Air Flow screen simply press the 12 Kestrel 4200 Pocket Air Flow Tracker Starting with a known altitude for your location You can obtain your altitude from a topographical map or local landmark Google Earth is an excellent free program that provides the exact altitude for any given address www earth google com Set this value as your reference altitude on the BARO screen to determine your baromet...

Page 4: ...DENSITY ALTITUDE screen is calculated from the absolute values of station pressure relative humidity and temperature and is not affected by the reference values entered in the BARO and ALTITUDE screens MAIN SETUP MENU You can customize your Kestrel Meter in multiple ways Press the button to access the Main Setup Menu Press the button to select the highlighted setting The Main Setup Menu contains O...

Page 5: ...ocket Air Flow Tracker APPLICATION EXAMPLES This section provides examples of applications where a Kestrel Meter might be used and the appropriate memory settings HVAC Environmental Monitoring Auto Store On Store Rate 5 min Overwrite On Man Store Off These settings will record conditions every five minutes for a total storage of almost 7 days You can monitor the conditions in a laboratory or manuf...

Page 6: ...s available for a more in depth definition On the internet visit www usatoday com or www noaa gov Or locate the USA Today publication The Weather Book Please note that any words in a definition printed in italics are themselves defined in this glossary 24 Kestrel 4200 Pocket Air Flow Tracker Some exercise physiology guides use wet bulb temperature rather than heat index as a measure of the safety ...

Page 7: ...ill be but for the most accurate readings at your particular location the Kestrel Meter is better CUSTOMER SERVICE Kestrel Pocket Weather Meters Warranty NK does not believe in disposable electronics We know that Kestrel Meters don t typically lead pampered lives and we design them for years of performance in tough conditions Every Kestrel Meter is designed and manufactured at NK s facility in Boo...

Page 8: ...s within 30 days of purchase and we will test and recalibrate all values at no charge Beyond 30 days we offer reasonably priced tests calibration services and N I S T certified calibrations as well as Kestrel Meter tune ups All of our measurements are traceable to the National Institute of Standards and Technology ensuring the highest level of accuracy Our primary Calibration Standards are sent fo...

Reviews: