The battery power is low.
Replace new batteries.
“LO” flashes 5 times before the
meter turning off automaticall
y.
The batteries are dead.
LO
LO
p
m
The environmental temperature
is too low to perform a test.
“LO” flashes 5 times.
Repeat the test in a war
m
place
about
57.2°F~104°F (14°C~40°C).
Y
ou may need to wait as long
as 20 minutes for the meter to
warm up before testing again.
The environmental temperature
is too high to perform a test.
“HI” flashes 5 times.
Repeat the test in a cool place
about
57.2°F~104°F (14°C~40°C).
Y
ou may need to wait as long
as 20 minutes for the meter to
cool down before testing again.
HI
p
m
Replace new batteries.
The meter starts to count down
then show your result on the
screen (10 seconds for glucose;
150 seconds for cholesterol
;
20 seconds for uric acid
;
10
seconds for hemoglobin
after the beep sound).
No action required.
Gl
u
p
m
p
m
CHO
L
p
m
UA
p
m
Hb
SYMBOL
WHAT IT MEANS
ACTION
41