![Kernel Multi Check User Manual Download Page 11](http://html1.mh-extra.com/html/kernel/multi-check/multi-check_user-manual_1957498011.webp)
1. Slide battery cover off from the back of the meter.
2. Insert two AAA size batteries (1.5V) into battery slot.
3. Slide battery cover back.
CAUTIONS:
◎
Make sure the “+” and “-“ ends of the batteries match the marks in the battery
compartment.
◎
Always replace both batteries at same time. Both batteries should be the same brand.
Using Check Strip to Check the Meter
You would need to use the check strip to check the meter when
You use
GCUHb Monitoring System for the first time.
You drop the meter.
You think there is something wrong with the results you got and you want to check whether
meter and test strips are working correctly or not.
1. Insert the check strip into the test strip slot on the meter.
2. The screen should display ”OK”. If “X” appears on the screen, remove the check
strip and reinsert it. If “X”displays again, stop the test and call our customer
support.
OK
11
Summary of Contents for Multi Check
Page 43: ......