9  Verification  

General hints: 
 
According to EU directive 90/384/EEC balances must be verified if they are used as 
follows (legally controlled area): 
 

a)  For commercial transactions if the price of goods is determined by weighing. 
b)  For the production of medicines in pharmacies as well as for analyses in the 

medical and pharmaceutical laboratory. 

c)  For official purposes. 
d)  For manufacturing final packages. 

 
In cases of doubt, please contact your local trade in standard. 
 
After verification the balance is sealed at the indicated positions. 
 

Verification notes: 

 

An EU type approval exists for the weighing system described in its technical data as 
verifiable. If the weighing system is used where obligation to verify exists as 
described above, it must verified and re-verified in regular intervals. 
Re-verification of a weighing system is carried out according to the respective 
national regulations. The validity for verification of weighing systems in Germany is 
e.g. 2 years.  
The legal regulation of the country where the balance is used must be observed! 
 

Check number:

 

After switching on the display unit the check number is displayed. This number will be 
automatically adjusted after every interference. In verified systems this number will 
be recorded by the weights and measures inspector. With that it is always possible to 
check whether a new adjustment has been carried out. 
In order to display the check number, switch off and switch on the system by  . The 
display unit shows subsequently e.g.: 
 

Ö

  8888.8 (check of LCD segments) 

Ö

  P 7.3 (software no.) 

Ö

  14325 (check number) 

Ö

  0 (weighing mode) 

 
 
 

VHM_M-BA-e-1012 

19

Summary of Contents for VHM_M

Page 1: ...egelei 1 D 72336 Balingen email info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instructions Pallet truck scales KERN VHM_M Version 1 2 07 2010 GB...

Page 2: ...sport 7 6 Unpacking Setup and Commissioning 7 6 1 Site of installation place of operation for scales 7 6 2 Scope of supply 9 6 3 Rechargeable battery operation 9 6 4 Commissioning 9 7 Display unit 10...

Page 3: ...cation value e 2 kg Verification class III Reproducibility 2 kg Linearity 2 kg Recommended adjustment weight not added class 2 t M1 Warm up time 10 min Allowable ambient temperature 10 C 40 C Humidity...

Page 4: ...Dimensions VHM_M BA e 1012 4...

Page 5: ...ung bezieht mit den nachstehenden Normen bereinstimmt Fran ais Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme aux normes cit es ci apr s...

Page 6: ...This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the balance minus a possibly existing tare load must be strictly avoided Balance may be damage by...

Page 7: ...he Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 4 2 Personnel training The appliance may only be operated an...

Page 8: ...for 0 1 of the weighed load is between 10 and 40 C Outside this temperature range deviations of up to 0 3 may occur Avoid jarring during weighing Protect the balance against high humidity vapours and...

Page 9: ...r a loss of rechargeable battery capacity is avoided When the system is used in shift work it is recommended to purchase a supplementary battery module The rechargeable battery is loaded using the del...

Page 10: ...ges may appear on the display HELP 1 Weighing system overloaded HELP 2 Taring impossible due to negative gross weight HELP 3 Negative signal of load cells to A D converter slant HELP 4 Entered tare we...

Page 11: ...e value in the flashing segment On Off Delete IMPORTANT The actuation of a key will only be accepted if the load is stable message load stable lit up The display unit only carries out functions as lon...

Page 12: ...absolutely clear In verified weighing systems observe the check number After switching on the display unit the check number is displayed This number will be automatically adjusted after every interfe...

Page 13: ...approx 18 seconds The display will disappear Keep pressed until the value of the first adjustment point appears The lower indicator W at the left edge of the display flashes Call up the three old adju...

Page 14: ...press the indicator bar flashes Press or until the highest adjustment point appears The upper indicator W flashes Press first digit is flashing Use or to set the flashing digit to zero Repeat this pr...

Page 15: ...the desired value will appear Confirm input by Change to the following digit using and if necessary change and acknowledge as described before Repeat this process for every digit After changing and co...

Page 16: ...e first digit flashes To change the flashing digit press or so many times until the desired value will appear Confirm input by Change to the following digit using and if necessary change and acknowled...

Page 17: ...ming the last digit the central indicator will flash For acknowledgement press 3 seconds The display counts reverse from AF08 to AF00 the second adjustment point is set Press the third adjustment poin...

Page 18: ...e display vanishes After adjustment the display device remains automatically in the higher resolution After switching off and on the factory side resolution will be displayed again If you readjust adj...

Page 19: ...ion to verify exists as described above it must verified and re verified in regular intervals Re verification of a weighing system is carried out according to the respective national regulations The v...

Page 20: ...y and a new verification wire seal mark fitted before it can be reused for applications subject to verification The verifiable model of the weighing system will display stripes only if a slant exceeds...

Page 21: ...resetting is herewith concluded the system returns automatically into the standard weighing mode 0 appears 10 2 Gross weighing After the load was lifted the display shows the gross value of the weighe...

Page 22: ...multiple range display in the standard version this are 200 kg see chap 10 1 To cancel the HELP display press and enter a new reduced tare weight Press button the most recently used tare value appear...

Page 23: ...e display unit offers an option that allows the addition of weighing values and the display of the total weight When the tare weight is active the net weight is automatically added up Add up Loading t...

Page 24: ...he following number 00 and the starting value 0 kg The system automatically reverts to standard weighing mode Please note that the system must be unloaded between the individual weighing procedures Ad...

Page 25: ...the manual lever should be left in central position This prolongs the life span of the seals The weighing system complies with the protection standard IP65 That means that dust and humidity such as ra...

Reviews: