8.2 Performance of adjustment
The weighing system off the possibility to enter one or maximally three adjustment
points (multi-point adjustment). This has the advantage that even weighing systems
showing a non-linear weighing curve may be easily adjusted within the required
limits.
1. Adjustment with one adjusting point
After zeroing (see chap. 8.1) the old adjustment points must be set to „zero“ as
follows.
¾
Start weighing system by pressing
.
¾
Press
approx. 18 seconds. The display will disappear. Keep pressed until
the value of the first adjustment point appears.
The lower indicator
W
at the left edge of the display flashes
¾
Call up the three old adjustment points using
(
Ð
) or
(
Ï
), whereby the
indicator at the left edge of the display moves upwards from the bottom.
When adjusting with only one adjustment point, the second and the highest
value must be set to „zero“ as follows.
¾
Press
(
Ð
) or
(
Ï
) until the second adjustment point appears. The
central indicator bar flashes.
¾
Press
, first digit is flashing.
¾
Use (
Ð
) or (
Ï
) to set the flashing digit to zero.
¾
Press
, the display changes to the following digit.
VHM_M-BA-e-1012
13