EFS-BA-e-2211
14
7.2 Menu overview
unit
Weighing Units: dwt, g, gn, oz, ozt, lb
cALE
Adjustment with specified adjustment weight
tr
Auto-Zero Tracking
AF
Auto off selectable after 30, 60, 180 sec., off
rESEt
Back to factory setting
EXit
Return to weighing mode
7.3 Description of individual menu items
7.3.1 Weighing units switch-over
Condition: Menu item „unit“
Confirm „unit“ by
, the weighing unit set as last is displayed.
Switch-over the weighing units by
until the desired unit appears. Confirm
these units by
, the unit flashes shortly before the balance returns to the
menu.
Press
to go till to the menu item „Exit“ and confirm by
. The balance
changes into weighing mode, the new weighing unit is now adjusted.
7.3.2 Zero tracking
The Auto-Zero function is used to tare small variations in weight automatically.
In the event that small quantities are removed or added to the material to be weighed,
incorrect weighing results can be displayed due to the “stability compensation“.
(Example: Slowly draining fluids from a container on the balance).
When apportioning involves small variations of weight, it is advisable to switch off this
function.
If
Zero-Tracking
however is switched off, the weighing display becomes more busy.