11  Verification  

General introduction: 
 
According to EU directive 90/384/EEC balances must be officially verified if they are 
used as follows (legally controlled area): 
 

a)  For commercial transactions if the price of goods is determined by weighing. 
b)  For the production of medicines in pharmacies as well as for analyses in the 

medical and pharmaceutical laboratory. 

c)  For official purposes. 
d)  For manufacturing final packages. 

 
In cases of doubt, please contact your local trade in standard. 
 
After verification the balance is sealed at the indicated positions. 

Verification of the balance is invalid without the "seal/lead seal". 

 
 
Verification instructions 

 

An EU type approval exists for balances described in their technical data as verify-
able. If a balance is used where obligation to verify exists as described above, it must 
officially verified and re-verified in regular intervals. 
Re-verification of a balance is carried out according to the respective national regula-
tions. The validity for verification of balances in Germany is e.g. 2 years.  
The legal regulation of the country where the balance is used must be observed! 
 

Balances with obligation to verify must be taken out of operation if: 

 

The 

weighing result 

of the balance is outside the

 error limit.

 Therefore, in 

regular intervals load balance with known test weight (ca. 1/3 of the max. load) 
and compare with displayed value.

 

 

The 

reverification deadline 

has been exceeded.

 

 

Summary of Contents for RXB

Page 1: ...Ziegelei 1 D 72336 Balingen E Mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instruction Price computing scale KERN RXB Version 1 2 11 2010 GB...

Page 2: ...the Operation Manual_______________ 13 5 2 Personnel training _______________________________________________ 13 6 Transport and storage __________________________________________ 13 6 1 Testing upon...

Page 3: ...0 11 Verification__________________________________________________ 22 11 1 Adjustment key and seals _________________________________________ 23 12 Service maintenance disposal_______________________...

Page 4: ...3 kg M2 3 kg M2 6 kg M2 6 kg M2 Net weight kg 4 4 4 8 4 4 4 8 300x330x110 without tripod 300x330x110 without tripod Dimensions of the housing B x D x H 300x330x110 300x330x540 with tripod 300x330x110...

Page 5: ...M2 15 kg M2 30 kg M2 30 kg M2 Net weight kg 4 4 4 8 4 4 4 8 300x330x110 without tripod 300x330x110 without tripod Dimensions of the housing B x D x H 300x330x110 300x330x540 with tripod 300x330x110 3...

Page 6: ...ere esta declaraci n est de acuerdo con las normas siguientes F D claration de conformit Nous d clarons avec cela responsabilit que le produit auquel se rapporte la pr sente d claration est conforme a...

Page 7: ...RXB BA e 1012 7 3 Appliance overview Models without vertical display Bubble level Secondary display on customer side Battery compartment Connection network cable ON OFF switch...

Page 8: ...RXB BA e 1012 8 Models with post mounted display Bubble level Post mounted display backlit revolvable Battery compartment Connection network cable ON OFF switch...

Page 9: ...RXB BA e 1012 9 3 1 Overview of display Operator display balance Weight Basic price 0 000 0 00 0 00 Loading status display Sales price Secondary display on customer page Weight Basic price Sales price...

Page 10: ...enter your basic price in kg or 100 g via keyboard The basic price can be switched to kg or 100 g Press key then 1 basic price in kg Press key then 2 basic price in 100 g The arrows next to the symbo...

Page 11: ...on 9 Number keys enter basic price PLU Enter 00 Press key then 1 basic price in kg Press key then 2 basic price in 100 g Delete total basic price Delete last input of basic price Zero key range 2 max...

Page 12: ...ate This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the balance minus a possibly existing tare load must be strictly avoided Balance may be dam ag...

Page 13: ...ety Precautions 5 1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if you are already familiar with KERN balances 5...

Page 14: ...t expose the device to extreme dampness for longer periods of time Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably war...

Page 15: ...Connect the power adaptor as soon as possible to change the battery The LED display provides information about the battery s charging status red Battery is almost discharged green Battery is complete...

Page 16: ...rice kg Price If you are using a weighing con tainer use the TARE key for taring 0 000 0 00 0 00 Place goods to be weighed on balance 1 300 0 00 0 00 Enter basic price via number keys sales price is a...

Page 17: ...aring 0 000 0 00 0 00 Place goods to be weighed on balance 1 300 0 00 0 00 Enter basic price via numeric keys sales price will be determined automatically and the basic price will be stored 1 300 5 00...

Page 18: ...ue 6 d or after key was pressed Display disappears after 5 sec when display moves towards zero Backlight off Press key CEnTR appears on the weight display Press the key during this display Backlight s...

Page 19: ...and keep on pressing it when balance is turned off Whilst doing this turn on balance Keep pressing down the key until the balance counts down to zero and the first function F0 appears Further function...

Page 20: ...settings are not allowed to be modified Press TARE key F8 Not documented settings are not allowed to be modified 10 3 Adjustment As the acceleration value due to gravity is not the same at every locat...

Page 21: ...objects on the weighing plate Operate menu activation key 10 1 stop check is carried out then the weight value of the calibrating weight will appear on the weight display Weight Basic price Sales pric...

Page 22: ...he seal lead seal Verification instructions An EU type approval exists for balances described in their technical data as verify able If a balance is used where obligation to verify exists as described...

Page 23: ...1012 23 11 1 Adjustment key and seals When verification of a scale is finished the positions indicated on the scale are sealed Verification of a scale without a seal is invalid Position of seals SEAL...

Page 24: ...d wipe with a dry soft cloth Loose residue sample powder can be removed carefully with a brush or manual vac uum cleaner Spilled weighing goods must be removed immediately 12 2 Service maintenance The...

Page 25: ...batteries inserted Draught air movement Table floor vibrations The displayed weight is permanently changing Weighing plate has contact with other objects Electromagnetic fields static charging choose...

Reviews: