7.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To re-
ceive accurate measuring values it is also recommended to adjust the balance peri-
odically in weighing operation.
Procedure when adjusting:
For verified balances adjustment by means of a jumper is locked. In order to carry
out adjustment the two contacts of the conductor plate must be shorted with a jumper
(see chapt. 7.7.1).
Observe stable environmental conditions. A warming up time of ca. 10 minutes is
required for stabilization. Ensure that there are no objects on the weighing plate.
Operation
Display
Switch on balance
PN
Use the number keys to enter password
„0000“:
confirmar
PN----
Ø
VnLQAD
Ensure that there are no objects on the
weighing plate.
The required adjustment weight is
displayed:
LQAD 6 HLOS
RPB-N-BA-e-0914
17