PLE-BA-e-0812
58
15 Error messages
Err2
Value outside zero range
Err3
Value outside taring range
Err4
Calibration weight outside allowable range (+-1% for calibration
weight)
Err5
Piece weight smaller than readability
Err7
Disconnection time was too short (should be more than 3
seconds)
Err8
Input outside permissible range
NULL
Impossible to carry out taring / resetting
FULL2
Weighing range exceeded
LH
Initial weight error Weight of weighing plate outside allowable
tolerance of 10%
5-FULL
Display range exceeded upon totalising
16 Service, maintenance, disposal
16.1 Cleaning
Before cleaning, please disconnect the appliance from the operating voltage.
Please do not use aggressive cleaning agents (solvents or similar agents), but a cloth
dampened with mild soap suds. Ensure that no liquid penetrates into the device and
wipe with a dry soft cloth.
Loose residue sample/powder can be removed carefully with a brush or manual
vacuum cleaner.
Spilled weighing goods must be removed immediately.
16.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized
by KERN.
Before opening, disconnect from power supply.
16.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to
valid national or regional law of the location where the appliance is used.