background image

21 

 

OZC-5-BA-e-1510 

Features 

 

 
 

Model outfit

 

Kern model 

Order  
number 

 

OZC 583 

 

Eyepieces 

HSWF 10x / Ø 23 mm 

●● 

OZB-A5503 

SWF 15x / Ø 17 mm 

○○ 

OZB-A5504 

SWF 20x / Ø 14 mm 

○○ 

OZB-A5505 

SWF 30x / Ø 9mm 

○○ 

OZB-A5506 

HSWF 10x / Ø 23 mm (with 0.1 mm scale) 

○ 

OZB-A5512 

SWF 15x / Ø 17 mm (with 0.05 mm scale) 

○ 

OZB-A5513 

SWF 20x / Ø 14 mm (with 0.05 mm scale) 

○ 

OZB-A5514 

C-mount 

0.3x 

○ 

OZB-A5701 

0.5x 

○ 

OZB-A5702 

1.0x 

○ 

OZB-A5703 

1.0x (with micrometre) 

○ 

OZB-A5704 

for SLR cameras (Nikon) 

○ 

OZB-A5706 

for SLR cameras (Olympus) 

○ 

OZB-A5707 

for SLR cameras (Canon) 

○ 

OZB-A5708 

Stand 

Arm curved, without lighting 

● 

 

Table

 

 

mechanical 

Dimensions WxD 188 x 160 mm,  
Travel:  76x65mm,  for  reflected  light  and 
transmitted light 

○ 

OZB-A5781 

Dimensions WxD 180 x 175 mm,  
Travel: 100x86mm, for reflected light 

○ 

OZB-A5782 

External

 

 

illumination 

For  information  on  external  lighting  units,  please  see  the  Kern  Optics  main 
catalogue, from page 68

 

and visit our website www.kern-sohn.com 

 

 

 

 

 

 

● = Standard configuration  

○ = Option

 

 

 

 

Summary of Contents for OZC 583

Page 1: ...RN Sohn GmbH Ziegelei 1 D 72336 Balingen E mail info kern sohn com Tel 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com User instructions Stereo zoom microscope KERN OZC 5 OZC 583 Ve...

Page 2: ......

Page 3: ...ality 11 5 1 Getting started 11 5 2 Adjust the interpupillary distance 11 5 3 Adjusting the magnification 11 5 4 Dioptre adjustment and focussing 12 5 5 Adjusting the stand 13 5 6 Using eye cups High...

Page 4: ...lements should be left as they were before use 1 2 Notes on the electrical system Before connecting to a mains power supply you must make sure that you are using the correct input voltage The informat...

Page 5: ...older must not be rotated back too far If you do this there is a risk that the microscope could tip over When the microscope is not being used you should fit the objective cap and cover the microscope...

Page 6: ...and alcohol as these are highly flammable substances You must therefore keep it away from naked flames and electrical devices which can be switched on and off and only use it in well ventilated rooms...

Page 7: ...OZC 5 BA e 1510 6 2 Nomenclature...

Page 8: ...7 OZC 5 BA e 1510 Side view of the microscope head...

Page 9: ...52 76 mm Dioptre adjustment On both sides Product dimensions WxDxH 305x180x405 mm Packing dimensions WxDxH 370x340x400 mm Gross weight 8 5 kg Net weight 6 6 kg Standard configuration Model KERN Tube E...

Page 10: ...r and head are then parallel on the central axis of the stand base see figure on page 10 Then you can remove the protective caps from the tube connectors so that you can then fit the eyepieces When do...

Page 11: ...OZC 5 BA e 1510 10 Assembled stereo zoom microscope...

Page 12: ...een the two eyepieces must be re adjusted While you are looking through the eyepieces use one hand to hold the righthand or lefthand prism housing firmly By rotating outwards or inwards you can either...

Page 13: ...ugh the right eyepiece with the right eye and bring the object into focus by using the focus control dials 5 Now set the largest possible zoom factor 6 Once again still only looking through the right...

Page 14: ...he other wheel Depending on the direction of the turn the torque will be increased or decreased On one hand this function can help to make it easier to adjust the focus and on the other hand it can pr...

Page 15: ...ed by the glasses symbol on the side They are also marked in the item description by an additional H example HSWF 10x 23 mm When fitting the eye cups make sure that the dioptre setting is not moved We...

Page 16: ...power adapter needed If the power connection is established you can turn on the incident lighting by using the on off switche of the control device see figure In addition the light intensity can be a...

Page 17: ...logen versions and also have an on off switch or different controller Typical goose neck lighting unit Using goose neck lighting units Depending on the requirements a goose neck lighting unit is place...

Page 18: ...this is on the top side of the microscope head see figure above When the trinocular toggle rod is pulled out the beam path which normally emits on the righthand eyepiece is diverted in the tube so tha...

Page 19: ...h an integrated micrometre in front of the camera C mount adapter You also need an adapter to connect a reflex camera But in contrast to the microscope camera this does not have a C mount thread but h...

Page 20: ...e right figure As the figures show both versions each have two coaxially positioned adjustment wheels in the righthand rear corner of the table This means that you can move it in the x and y direction...

Page 21: ...ulbs Due to the long service life of an LED lighting system for these microscopes it will not be necessary to simply change a bulb Problems with the lighting unit would therefore in most cases be caus...

Page 22: ...x OZB A5703 1 0x with micrometre OZB A5704 for SLR cameras Nikon OZB A5706 for SLR cameras Olympus OZB A5707 for SLR cameras Canon OZB A5708 Stand Arm curved without lighting Table mechanical Dimensio...

Page 23: ...ufficient The wrong bulb has been used The input voltage is too low Optical unit Problem Possible causes You can see two images The interpupillary distance is not set correctly The magnifications of t...

Page 24: ...materials which you can dispose of at your local recycling centre Disposal of the storage box and device must be carried out by the operator in accordance with all national or regional laws in force...

Page 25: ..._____________________________________________________ __________________________________________________________ __________________________________________________________ ____________________________...

Reviews: