
14
MWB-SH-e-1810.docx
9 Error Messages
Display
Description
Zero range exceeded
(on start-up or when pressing the
key)
Load on weighing pan
Excess load, during zero setting of weighing scale
Incorrect adjusting process
Fault on load cell
Value outside the A/D converter range
Damaged weighing cell
Damaged electronics
Unable to initialize zero point
Measuring cell defective / overloaded
Object on weighing pan / contact
Transport safety device has not been removed
Main board defective
OL or --------
Overload
Unload, switch off and adjust the balance
------
or „0“
Underload
Unload, switch off and adjust the balance
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.