MWA-M / MWA-PM-BA-e-1614
37
13.3 Printer operation
Printout examples:
Prt
0 / 2
60.0kg
1 / 3
60.0kg
170.0cm
20.7BMI
14 Error messages
Display
Description
Zero range exceeded
(on start-up or when pressing the
key)
•
Load on weighing pan
•
Excess load, during zero setting of weighing scale
•
Incorrect adjusting process
•
Fault on load cell
Value outside the A/D converter range
•
Damaged weighing cell
•
Damaged electronics
Should other error messages occur, switch balance off and then on again. If the error