MSF-BA-EU-1818 

 

Unpacking/installing 

Testing upon 

acceptance 

Please check packaging as well as the device on receipt for possibly visible 
external damages. 

Packaging/ 

Return 

Keep all parts of the original packaging for a possibly required return.  
Only use original packaging for returning. 

Checking the 

graduation 

After the assembly of the height measuring rod, the graduation must be 
checked using an external measuring rod. 
For this put an external, verified measuring rod with the respective accu-
rateness on the weighing plate or, in case of wall assembly, vertically on 
the floor. The height measuring rod must be shifted in a way that the hori-
zontally unfolded flap touches the external measuring rod with the lower 
side at 150 cm. 
Information:   
The measure of the external measuring rod is not congruent with the grad-
uation of the height measuring rod, it is displaced by some cm, as the offset 
between the flap and the reading mark thus will be considered. 
If the measure does not match exactly, loosen again slightly and shift a little 
the screws used for assembly and then check again the measure of 150 
cm. 

 

Scope of supply

 

 

2 x bracket 

 

2 x screw „Measuring rod on tripod of 

scales“ 

 

4 x dowel, “wall bracket“ 

 

4 x screws „Measuring rod on wall with 

bracket“ 

 

4 x screw, “scales stand 

bracket“ 

 

Operating instructions 

 
 

 