![KERN KERN ALT Operating Instructions Manual Download Page 52](http://html1.mh-extra.com/html/kern/kern-alt/kern-alt_operating-instructions-manual_1957295052.webp)
ALT-BA-defsi-0410
120
8.1.5 Percent determination
Percent determination allows weight display in percent, in relation to a
reference weight
Display overview for percent determination operation:
1. % - deviation to reference weight
2. Weighing-in aid coarse/fine (only displayed for active dispensing)
3. Mode of operation
4. Present weight of sample
5. Reference weight
6. Tolerance marker (only displayed for active tolerance weighing)
7. Percent determination mode
Operator
indication
Ö
Call up operating mode
“M4 Percent“
(see chapter 8.1)
Ö
Press -key
01 Reference – Numeric entry of the reference weight
Ö
Use the cursor (
►
) to select “
01 Reference“
Ö
Press the -key; the current menu item blinks.
Use the arrow keys (see chapter 7.1) to enter reference
weight and confirm with the
-key.
Ö
Press the -key; the balance is now in percent
determination mode.
Ö
Now place items to test onto the weighing plate, the
percentage in relation to the reference part is displayed.