Now place the adjusting weight in the center of
the weighing plate, press the
-key. After
successful adjustment the system will
automatically return to the menu.
The
-key can be used to cancel the
adjustment.
In case of an adjustment error or incorrect
adjusting weight an error message is displayed.
Repeat adjustment.
Setup
P 1 01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
fun c tio n
fun c tio n
fun c tio n
o ff
o n
fun c tio n
31.01.05
1 1 h o u r
We recommend:
Carry out adjustment as close to the maximum load of the balance as possible.
Information concerning the test weights is available at:
http://www.kern-sohn.com
04 Calibration test
Here, deviation from the last adjustment is determined. This is only a check, i.e.
no values are changed.
Ö
Use the cursor (
►
) to select
“04 Cal. test“
Setup
P 1 01 Int. calibr.
02 Ext. calibr.
03 User calibr.
04 Calibr. test
05 Weight. corr.
06 Aut calibr.
07
Print report
o
Auto cal. time
08
********
********
********
********
0.0
0
1
fun c tio n
fun c tio n
fun c tio n
o ff
o n
fun c tio n
31.01.05
1 1 h o u r
Ö
Press the
-key, the calibration test is started
Calibration
Calibration check
Ö
The result is displayed
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
05 Weight corr. – not documented