6  Unpacking and implantation 

6.1  Installation Site, Location of Use 

The balances are designed in a way that reliable weighing results are achieved in 
common conditions of use.  
You will work accurately and fast, if you select the right location for your balance. 
 

On the installation site observe the following: 

  Place the balance on a firm, level surface; 

  Avoid extreme heat as well as temperature fluctuation caused by installing 

next to a radiator or in the direct sunlight; 

  Protect the balance against direct draughts due to open windows and doors; 

  Avoid jarring during weighing; 

  Protect the balance against high humidity, vapors and dust; 

  Do not expose the balance to strong humidity for extended periods. Non-

permitted condensation (condensation of air humidity on the appliance) may 
occur if a cold appliance is taken to a considerably warmer environment. In 
this case, acclimatize the disconnected appliance for ca. 2 hours at room 
temperature. 

  Avoid static charge of goods to be weighed or weighing container. 

  Being of protection type IP 67 as per DIN EN 60529, the weighing scale is 

suitable for short-term use in wet conditions.  

 
Major display deviations (incorrect weighing results) may be experienced should 
electromagnetic fields (e.g. due to mobile phones or radio equipment), static 
electricity accumulations or instable power supply occur. Change location or remove 
source of interference.  
 

6.2  Unpacking/implantation 

Take the balance carefully out of the package, remove the plastic cover and  

place it in the designated work area. 

Scope of delivery / serial accessories: 

 

  Balance, see chpt. 2 

  Mains power supply

 

 Operating 

instructions

 

 
 

Summary of Contents for IFB-BA-e-1012

Page 1: ...bH Ziegelei 1 D 72336 Balingen email info kern sohn com Phone 49 0 7433 9933 0 Fax 49 0 7433 9933 149 Internet www kern sohn com Operating instructions Platform scale KERN IFB Version 1 2 02 2010 GB I...

Page 2: ...10 3 2 Improper Use 10 3 3 Warranty 10 3 4 Monitoring of Test Resources 10 4 Basic Safety Precautions 11 4 1 Pay attention to the instructions in the Operation Manual 11 4 2 Personnel training 11 5 Tr...

Page 3: ...utomatic adding up 22 7 9 Animal weighing 23 7 10 Lock keyboard 23 7 11 Display background illumination 23 7 12 Automatic switch off function AUTO OFF 24 8 Menu 25 9 Service maintenance disposal 27 9...

Page 4: ...1 g 2 g 2 g 5 g 2 g 5 g Recommended adjustment weight not added class 6 kg M1 15 kg M1 15 kg M1 Warm up time 10 minutes Stabilization time typical 2 sec Weighing unit kg Auto Off Available options 5 1...

Page 5: ...g 10 g 10 g 20 g 10 g 20 g Recommended adjustment weight not added class 30 kg M1 60 kg M1 60 kg M1 Warm up time 10 minutes Stabilization time typical 2 s Weighing unit kg Auto Off Available options...

Page 6: ...ity 20 g 50 g 20 g 50 g 50 g 100 g Recommended adjustment weight not added class 150 kg M1 150 kg M1 300 kg M1 Warm up time 10 minutes Stabilization time typical 2 s Weighing unit kg Auto Off Availabl...

Page 7: ...2 Appliance overview 1 Battery status display 2 Keyboard 3 Weight display 4 Tolerance tag see chpt 7 6 5 Weighing unit 6 Levelling screw 7 Spirit level underneath weighing platform IFB BA e 1012 7...

Page 8: ...Display sum total Navigation key Digit selection to the right Add weighing value in summation memory Navigation key Digit selection to the left Calculate weighing data via interface C Delete Switch ov...

Page 9: ...st digit after the last digit has been pressed To change the selected flashing digit press repeatedly until the desired value appears Then select by using additional digits and change these by using F...

Page 10: ...on protected The structure of the balance may not be modified This may lead to incorrect weighing results safety related faults and destruction of the balance The balance may only be used according to...

Page 11: ...upon acceptance When receiving the appliance please check packaging immediately and the appliance itself when unpacking for possible visible damage 5 2 Packaging return transport Keep all parts of th...

Page 12: ...on permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimatize the disconnected appliance...

Page 13: ...ion only if the balance has not already been adjusted to the location in the factory This adjustment process must be carried out for the first commissioning after each change of location as well as in...

Page 14: ...re no objects on the weighing plate The currently set adjustment weight will be displayed Change by pressing the navigation keys See chpt 2 1 1 select desired setting currently enabled digit will be f...

Page 15: ...ualification Approval is in existence for verified weighing systems If a balance is used where obligation to verify exists as described above it must officially verified and re verified in regular int...

Page 16: ...can be reset to zero whenever needed as described below Remove load from weighing system Press and the zero display as well as the ZERO indicator will appear 7 4 Simple weighing Place goods to be wei...

Page 17: ...between gross weight and net weight by pressing the key To delete the tare value remove load from weighing plate and press 7 6 Weighing with tolerance range You may determine an upper and lower limit...

Page 18: ...ce limit Red signal lamp glowing Settings for tolerance weighing may be set either by calling up menu block P0 CHK See chpt 8 or by applying the faster option of pressing the key combination Settings...

Page 19: ...e no ok ng to select the desired setting Confirm input by Press and the weighing system will go into tolerance weighing mode From here evaluation takes place whether the goods to be weighed are within...

Page 20: ...goods can only be added when the display zero Place goods to be weighed B Wait until the stability display appears then press The weight value will be added to the summation memory and possibly print...

Page 21: ...e same time The data in the summation memory are deleted IFB BA e 1012 21 1 2 3 4 Printout example 1 First weighing 2 Second weighing 3 Third weighing 4 Number of weighings total NO 1 1 000kg NO 2 0 5...

Page 22: ...rinting Remove the weighed good More weighed goods can only be added when the display zero Place goods to be weighed B After the standstill control sounds a signal tone The weighing value is added to...

Page 23: ...animal weighing function is enabled During the calculation of a mean average you can add or remove loads as the mean average will be continuously updated To disable the animal weighing function press...

Page 24: ...unction AUTO OFF The instrument will switch off automatically after a set time when the display unit or weighing bridge has been idle Keep pressed 3s until setbl appears Call up AUTO OFF function usin...

Page 25: ...f the individual menu items can be selected one after the other Select setting Confirm selected menu item with The current setting will be displayed Change settings Use the navigation keys see chpt 2...

Page 26: ...mory and issued AUTO Automatic add up see chpt 7 8 This function is used to automatically add and send individual weighing values to the summation memory whilst weighing instrument is being unloaded M...

Page 27: ...onal or regional law of the location where the appliance is used 9 4 Error messages Error message Description Possible causes Maximum load exceeded Remove load from scale or reduce preload Zeroing ran...

Page 28: ...suitable cable to connect the weighing balance to the interface of the printer Faultless operation requires an adequate KERN interface cable Communication parameters baud rate bits and parity of balan...

Page 29: ...eight NT Net weight lf Space line lf Space line Printout summation memory lf Space line TOTAL NO 3 Number weighing processes TOTAL wgt 0 447KG Total of all individual weighings 10 3 Continuous data ou...

Page 30: ...ted The displayed weight is permanently changing Draught air movement Table floor vibrations Weighing plate has contact with other objects Electromagnetic fields static charging choose different locat...

Reviews: