7. Adjustment
Switch-off balance and attach a carrying help if necessary.
Switch-on balance with attached carrying help and during the selftest press the
ZERO
and the
button at the same time. „
P1 - - -
“is displayed.
Use the number keys to enter password:
or
select standard password „0000“ with
HOLD
number, increase the flashing
number using
or
personal password, see function F8 ci
Confirm with
ZERO
button, the first function „F0 di“ is displayed.
Press the
button repeatedly until „
F9 CAL
“ will be displayed.
Press the
ZERO
button, „UnLD“ will be displayed.
Unload the balance and wait until the LED
lights up.
Press the
ZERO
button, the currently set adjustment weight is displayed.
In order to change, select the number to be changed using the
HOLD
button and
set the desired value using the
button, the respective active digit
flashes.
Confirm by the
ZERO
button, „Load“ will be displayed.
Attach the adjustment weight and wait until the LED
lights up.
Press the
ZERO
button.
After successful adjustment the balance carries out a selftest, then it
automatically returns to weighing mode.
An adjusting error or incorrect adjusting weight will be indicated by the error
message; repeat adjustment procedure
Summary of Contents for HFM 10T1
Page 33: ...HFM BA d 1813 32...
Page 36: ...HFM BA d 1813 35 Kranwaage Modell Serien Nr Teil Ma nahme Datum Name Unterschrift...
Page 37: ......
Page 67: ...HFM BA e 1813 31...
Page 70: ...HFM BA e 1813 34 Crane scales Model Serial no Part Action Date Name Signature...
Page 71: ......
Page 101: ...HFM BA f 1813 31...
Page 104: ...HFM BA f 1813 34 Balance grue Mod le N de s rie Pi ce constitutive Mesure Date Nom Signature...