HDB-BA-e-1730
25
6.7.1
Data-Hold function
Switch on the balance, keep the
HOLD
-button pressed until the current setting
„Hx“ (H1 – H6) appears.
Press
ON/OFF
button repeatedly until the desired setting
„H1-H4“
is displayed.
Confirm setting with the
HOLD
button.
Suspend the material to be weighed
The weight value is fixed and displayed depending on the setting (H1
– H4) (see
chap. 5.7), represented by the [
] in the upper left.
6.7.2
Animal weighing function
This function is suitable for busy weighing procedures. The result is a mean value
formed by 16 weighing values which is found out within 3 seconds.
Switch on the balance, keep the
HOLD
-button pressed until the current setting
„Hx“ (H1 – H6) appears.
Press
ON/OFF
button repeatedly until the setting
„H5“
is displayed.
Confirm setting with the
HOLD
button.
Suspend the material to be weighed
Press the
HOLD
-button, the display counts reverse from 3 -1.
The calculated mean value is indicated, symbolised by the [
] top left.
Before further measurements press first the
ON/OFF/TARE
-button.