9.4 Linearization (non-verified models only)
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.
•
Carrying out linearization is restricted to specialist staff possessing
well acquainted with the workings of weighing scales.
•
The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter 2.4 “Testing instruments control”.
•
Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
•
After successful linearization you will have to carry out calibration;
see chapter 2.4 “Testing instruments control”
Tab. 1: Adjustment points
Adjustment
weight
GAB 6K0.05N
GAB 12K0.1N
GAB 30K0.2N
1.
0 kg
0 kg
0 kg
2.
2 kg
4 kg
10 kg
3.
4 kg
8 kg
20 kg
4.
6 kg
12 kg
30 kg