Cautions
(1) [ ] must be carried out in following cases:
(1)
When built-in weight is added or replaced.
(2)
When linearity adjustment is done.
(3)
When AJDP board is replaced.
(2) The quality/tolerance of the reference weight determine the accuracy of the scale.
Use weights of higher accuracy than the scale.
(3) Adjust the level beforehand. The REF CAL must be done in a good environment,
no wind, no oscillation, and no temperature changes.
(4) : The reference weight is less than ½ of F.S.
: The data error exceeds 1%. Adjust the linearity.
- 24 - EW_EW-C_EG_N-SH-e-0722
Summary of Contents for EG
Page 6: ...2 Electronic Construction 2 1 Block Diagram EG EG AC Adaptor 6 EW_EW C_EG_N SH e 0722...
Page 7: ...2 2 Whole Wiring EG EG 7 EW_EW C_EG_N SH e 0722...
Page 8: ...3 Troubleshooting 3 1 Troubleshooting Procedure 8 EW_EW C_EG_N SH e 0722...
Page 13: ...4 1 2 EG Type 13 EW_EW C_EG_N SH e 0722...
Page 16: ...4 3 2 EW 2200 12K 16 EW_EW C_EG_N SH e 0722...
Page 17: ...4 4 Adjustment Sequence For Cases 17 EW_EW C_EG_N SH e 0722...
Page 19: ...4 6 How To Call Address Mode How To Re write Address Data EW C Type 19 EW_EW C_EG_N SH e 0722...
Page 20: ...4 7 Linearity Adjustment 1 How To Adjust Linearity EW C Type EG type 20 EW_EW C_EG_N SH e 0722...
Page 28: ...5 1 4 EW 120 4 NM 28 EW_EW C_EG_N SH e 0722...
Page 30: ...5 3 Sequence Of Mechanism Unit Replacement 30 EW_EW C_EG_N SH e 0722...
Page 34: ...5 4 Plate Rib Replacement 34 EW_EW C_EG_N SH e 0722...
Page 45: ...7 2 EW 2200 12K 45 EW_EW C_EG_N SH e 0722...
Page 46: ...7 3 EW C EW 120 4 NM EW 600 C3 NM EW 120 4 NM 46 EW_EW C_EG_N SH e 0722...