31
SDBY
Auto Sleep
function
SDBY Y
Function enabled
The scale is switched to “Sleep” mode automatically
after 50 seconds of no operation or load change. The
indication will disappear, solely the dot will be lit.
Once the load is changed or the scale is operated,
the scale is switched on again automatically.
SDBY N
Function disabled
FiLt-1
Filter
FiLt-1*
∼
FiLt-5
Adaptation to the
ambient conditions
, you can se-
lect from FiLt-1
∼
FiLt-5.
The higher the filter degree, the faster the response
time/the higher the sensitivity.
Zero-1
Maintaining zero
ZEro0*
∼
ZEro9
Automatic zero maintenance, possible to choose
from 0
d
to 9
d
If the amount of the weighed material is reduced or increased
significantly, the scale’s “stabilizing and compensating” mecha-
nism can result in displaying erroneous weighing results! (e.g.:
slow outflow of the liquid from the container placed on the scale,
evaporating processes).
When dosing with small weight fluctuations, it is recommended
to switch this function off.
L-AZ-0
Setting
a decimal point
L-AZ 0*
∼
L-AZ 9
The load scope where the scale returns to zero, you can
choose from 0
d
– 9
d
ledb
LED
brightness
ledb-h
bright
ledb-m
medium
ledb-l
low