ABT-BA-e-1312
45
12.8 Settings control display
To receive a confirmation of the current settings, repeatedly actuate the
[CAL]
key in
weighing mode until "St. b1 t" (ex.) appears.
Stability and response setting
St:
Standard mode
Co: Anti-convection mode
Hi:
High-stability mode
Po: Apportioning mode
Stability and response setting
b1: 1 count
b5: 5 count
b10: 10 count
Appears when zero
function is turned on.
The weight symbol is displayed when either automatic adjustment PSC or Clock-CAL
or both functions are turned on.
Appears when the automatic adjustment has
been switched on by PSC or Clock-CAL.
Appears when the issue of the adjustment
data is switched on.