572/573/KB/DS-BA-defsi-04410
36
JUSTIEREN/ADJUST
The balance must be
adjusted
at its place of installation before initial use and at regular
intervals.
Please observe the warm-up time referred to in the chapter on ”Commissioning”.
It is essential to avoid jarring and disturbances during the adjusting procedure.
TARAVORABZUG/PRESELECT TARE
The known tare value of a weighing container can be “deducted” by entering its weight as
preselect tare
. In this way only the net weight of the goods to be weighed is displayed during subse-
quent weighings. PTA will be seen on the display.
Manual tare may not be carried out using the TARE key.
AUTO TARE ein/active
AUTO TARE aus/inactive
Auto tare
activation serves to stabilize the zero point of the balance. Minor changes in weight in the
zero point range are tared automatically, i.e. the displayed figure remains at zero.
GESCHWINDIGKEIT/Filter/SPEED
The balance can be adapted in stages to its location from 1 – 5.
Stage 1
= good installation conditions,
fast display
/
low filter effect
(e.g. proportioning)
Stage 5 = poor installation conditions
, slow display / high filter effect
(restless environment)
e.g. proportion weighing requires a greater display speed and this can be adjusted by selecting ”Fast"
in the MODE program.
AUTO OFF ein/active
AUTO OFF aus/inactive
The
auto-off
function inactivates the balance after approx. 60 seconds if it is not in use.
VARIABLER FACTOR/VARIABLE FACTOR
The weighing value in grams is automatically multiplied by the set
variable factor
and the result (in-
cluding the unit *) shown on the display.
Example:
A sheet of paper sized 10 x 10cm weighs 0,6 g. In order to determine the weight g/1m², the factor has
to be at 100. The display value is 0,6g x 100 = 60,0*, thus 60,0 g/m².
VOREINSTELLUNG/PRESELECT UNIT
UNIT-Auswahl/select
All selected units checked YES in
Preselect Units
are offered on the UNIT key in weighing mode
for unit changeover.
Recommendation: Only preselect the units that are actually required.
Each time the
UNIT
key is pressed, the next preselected unit
(using Preselect Units)
is selected.
Summary of Contents for 572 DS
Page 12: ...572 573 KB DS BA defsi 04410 35 ...