4. Set the output on or off as desired using the

STANDBY

 key. The output is off (disabled) when

the 

STANDBY

 indicator is lit, on (enabled) when

not lit. 

5. Use 

Y

 or 

U

 to highlight the main channel. There

are two ways to program the output in local mode.
These methods can be used either when the out-
put is disabled (

STANDBY

 indicator lit) or

enabled.

WARNING

When the ADJUST control is
rotated, the active parameter is
immediately effective if the output
is enabled (on = STANDBY indica-
tor not lit). The voltage/current
applied to the load changes as the
ADJUST control is rotated. 

Use the 

ADJUST

 control to increase or

decrease the main channel setting (e.g., volt-
age when the unit is in voltage mode). Start
with the most significant digit of the desired
value, then use 

T

 to highlight the next digit.

For fine adjustment press the 

ADJUST

 control

in while rotating the knob to modify the least
significant digit. 

Enter the desired value on the keypad using
the number keys. For example, to program the
BOP to 75.8V, press the following keys in
order 

75.8

 then press 

ENTER

. For fine

adjustment use 

Y

, and 

U

 to modify the least

significant digit. To correct the entry before
activation press the 

CLEAR

 key to set the

value to zero and start over. When the desired
value is displayed, press 

ENTER

. This causes

the new value to appear at the output and be
applied to the load if the output is enabled.

6. To program the corresponding Protect channel,

press 

Y

 or 

U

 as necessary to highlight the Pro-

tect channel. Then set the value using either of
the two methods described above. If the Protect
Entry setting is set to Independent, separate
entries for the positive and negative protect chan-
nel are possible. Otherwise the value entered is
applied to both positive and negative protect
channels.

NOTE: The BOP can be configured to show

the protection limits as either a single
value that applies to both protection
channels or show individual settings
for positive and negative protection
limits. See Operator’s Manual for
details

SOFTWARE LIMITS.  

Software limits prevent pro-

gramming of the main channel or the Protect channel
beyond the software limit value. Refer to Operator’s
Manual for a full explanation of software limits.

Changing Main Channel Software limit. 

This procedure allows the user to determine the maxi-
mum value of voltage or current that can be pro-
grammed.

1. Press 

%

 from the power-up screen to enter the

General Setup menu, then highlight Max/Min Set-
tings. 

2. Press 

!

 to enter the Max/Min Settings submenu

(Figure 8). (If a Password is required, see Opera-
tor’s Manual for instructions.) 

3. Highlight the voltage or current max/min value

and press 

!

 to change it. Software limits are

absolute values (do not use minus sign for nega-
tive limits). Use number keys to change the set-
ting, then 

$

 to save. 

4. When complete, press 

$

 to save for power-up,

#

 to abort, or 

%

 to apply the changes (without

saving for power-up) and exit.

5. Upon return to the power-up screen, the main

channel (voltage or current) is compared against
the main channel limits in effect. If the main chan-
nel exceeds the limit, it is set to zero. 

6.  Highlight the ±CPROTECT or ±VPROTECT max/

min value and press 

!

 to change it. Software

limits are absolute values (do not use minus sign
for negative limits). Use number keys to change
the setting. Press 

$

 to save, or 

%

 to abort. 

7. When complete, press 

$

 to save for power-up,

or 

#

 to abort, or 

%

 to apply the changes (with-

out saving for power-up) and exit.

Upon return to the power-up screen, the new protection
limit (voltage or current) is compared against the pro-
tection limits in effect. If the new protection limit setting
is below the existing setting for the protection limit, the
protection channel (voltage or current) is set to zero.

Summary of Contents for BOP 10-200ME

Page 1: ...ing instructions including an Installation Operation Summary which includes hyperlinked references to detailed procedures but which can be printed as a handy reference The BOP 2KW MG ME Operator s Man...

Page 2: ...ering many electrochemical reactions Models with L suffix have been optimized for exceptionally low current ripple and noise and improved stability drift and temperature making them ideal for driving...

Page 3: ...th displayed voltage on LCD within 0 05 of rated maximum voltage VI ACCESSORIES See Table 3 TABLE 3 ACCESSORIES ITEM FUNCTION PART NUMBER Mating Connector Trigger Mates with Trigger port 142 0527 Kepc...

Page 4: ...4 228 1915 053119 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 1 BOP OUTPUT CHARACTERISTICS...

Page 5: ...Terminal Block Then connect OUT S and COM S lines from the BOP 2KW Sense Monitor Terminal Block to the load see Figure 7B using 22 AWG wire twisted pair ANALOG I O CONNECTIONS The Analog I O Port conn...

Page 6: ...load exists contacts cannot be held closed by actuator When the power supply is turned on it per forms a brief self test that includes testing the three processors analog interface and dis play and an...

Page 7: ...con figuration shown as a power up setting so the changes will not be lost when the unit is turned off APPLY EXIT applies the current changed setting without saving for power up and exits to the previ...

Page 8: ...AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 BOP 2KW SERIES POWER BOOST SECTION FRONT PANEL CONTROLS AND INDICATORS FIGURE...

Page 9: ...o incre ment or decrement the highlighted digit Pressing ADJUST while rotating adjusts the least significant digit If the output is on unit not in Standby changes made using the ADJUST control are imm...

Page 10: ...channel is determined by the MODE key in local mode which alternately selects either Voltage mode or Current mode or by SCPI com mand in digital remote mode The protection channel is determined automa...

Page 11: ...l are possible Otherwise the value entered is applied to both positive and negative protect channels NOTE The BOP can be configured to show the protection limits as either a single value that applies...

Page 12: ...loads Table 4 indicates how the power supply responds to a command to go from Output ON to OFF When the output is disabled the unit is set to voltage mode voltage is set to zero and both current prot...

Page 13: ...opened after the connections are complete and then the switch in series with the battery should be closed If the constant voltage type active electronic load is adjusted to zero before handling the p...

Page 14: ...com mands via RS 232 GPIB or LAN ports opera tion via web pages Analog Remote Operation via Analog I O port Details about Protect Limits and Software controlled limits Storing Recalling Power Supply...

Page 15: ...15 15 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 9 BOP 2KW OUTLINE DIMENSIONS Sheet 1 of 2 REAR VI...

Page 16: ...NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 10 BOP 2KW WIRING DIAGRAM DETAIL A S MON OUT OUT MON COM GND GND NET OUT MON N C OUT S COM S CO...

Reviews: