Kenwood PS10-35 Manual Download Page 5

Power source

85 to 132V /170 to 250V
AC 50 to 60Hz

Display instruments

Voltmeter

Display

3-1/2 digits red LED

Accuracy 0.1

rdg

±

2digits (23

±

5

)

Temperature coefficient

±

100ppm/

(0 to 50

)

Ammeter

Display

3-1/2 digits red LED

Accuracy 0.5

rdg

±

3digits (23

±

5

)

Temperature coefficient

±

200ppm/

(0 to 50

)

Protection

Overvoltage protection

The power relay is tripped when the
output voltage is exceeded by approx.
110%. of the rated voltage.

Overcurrent protection

The power relay is tripped when the
output cur rent is approx. 110% to
130% of the rated current.

Overheating Protection

The power relay is tripped when the
temperature of the heat sink is
approx. 100

±

5

°

C or more.

Thermal fuse

Built-in thermal fuse

Input fuse

Cut off the power relay

REGULATED DC POWER SUPPLIES

62

Common Specifications

Environmental condition

Temperature/humidity
for operation

0 to 50

/30 to 80

RH

Temperature/humidity

for storage

20 to 70

/20 to 80

RH

Cooling system

Forced air (fan), front air intake type.

Functions

Automatic output switch resetting,
voltage/current limit switch, preset
OVP

Case dimensions

Type I

70 (W)

× 

124 (H)

× 

350.5 (D) mm

Type II 

140 (W)

× 

124 (H)

× 

350.5 (D) mm

Type III

210 (W)

× 

124 (H)

× 

350.5 (D) mm

Accessories

2.5m AC cable 

× 

1

SPECIFICATIONS

Output

CV (constant-voltage) characteristic

Model

CV

CC

Ripple*

Line regulation Load regulation

Transient

Rise time/

response**

Fall time

V

A

mV rms

0.05%

mV

0.1%

mV

m sec

m sec (Full loaded)

PS6-60

0 to 6

0 to 60

10

5

5

1

80/150

PS6-120

0 to 120

2

PS10-35

0 to 35

1

PS10-70

0 to 10

0 to 70

10

5

5

2

80/150

PS20-18

0 to 18

10

1

PS20-36

0 to 20

0 to 36

5

5

80/150

PS20-54

0 to 54

15

2

PS36-10

0 to 10

1

PS36-20

0 to 36

0 to 20

10

5

5

2

80/150

PS36-30

0 to 30

15

PS60-6

0 to 6

10

1

PS60-12

0 to 60

0 to 12

15

5

5

80/150

PS60-18

0 to 18

20

2

Summary of Contents for PS10-35

Page 1: ...s are also provided including remote control functions with GP IB control capability which is possible when the optional GP 600 is used Compact Size Light Weight High Efficiency Thanks to the advanced switching method the PS Series power supplies are both compact and light with less than half the mass and volume of previous power supply models resulting in an excellent space factor and high effici...

Page 2: ...ircuit is activated or when the power is switched off an alarm signal is generated to turn the output off and interrupt the AC power relay Power relay off signal When trouble occurs on the user side etc the power relay is interrupted by an external signal Rush current protection circuit The rush current protection circuit is activated at the moment of the power is switched on to reduce the input s...

Page 3: ...for remote control CV CONTROL Control by an external voltage SELECT Control by an external resistance Normal mode S1 Control by an external resistance Fail safe mode Remote local switching of output current voltage resistance mode selection for remote control CC CONTROL Control by an external voltage SELECT Control by an external resistance Normal mode Control by an external resistance Fail safe m...

Page 4: ...ector Output voltage Output voltage 0 to 10V 0 to10kΩ to 0Ω S terminal J1 Output current Output current 0 to 10V 0 to10kΩ to 0Ω S terminal J1 Output ON OFF Output ON Photo diode ON Floating Community common J1 Power ON OFF AC power relay ON Photo diode ON Floating Community common J1 CV mode signal At CV operation Photo transistor ON Floating Community common J1 CC mode signal At CC operation Phot...

Page 5: ...ifications Environmental condition Temperature humidity for operation 0 to 50 30 to 80 RH Temperature humidity for storage 20 to 70 20 to 80 RH Cooling system Forced air fan front air intake type Functions Automatic output switch resetting voltage current limit switch preset OVP Case dimensions Type I 70 W 124 H 350 5 D mm Type II 140 W 124 H 350 5 D mm Type III 210 W 124 H 350 5 D mm Accessories ...

Page 6: ...e rated voltage Figure of Case dimensions 1 RB 600A 1 2 rack width RB 600B 1 3 rack width RB 600C 1 6 rack width CC constant current characteristic Others Ripple Line regulation Load regulation Input Current Case dimensions Weight mA rms 0 2 mA 0 2 mA AC 100 200V A Type kg approx 120 5 5 8 5 I 3 0 260 10 10 16 10 II 5 5 70 5 5 8 5 I 3 0 160 10 10 15 9 II 5 5 40 5 5 8 5 I 3 0 92 10 10 15 9 II 5 5 1...

Reviews: