Procedure for Use
1. Place the anesthetized animal on the warming pad directly on top of the Pad
Sensor.
2. If using Homeothermic regulation, insert the animal temperature probe into
the rectum of the animal. Secure the probe if necessary.
Always secure the pad temperature probe beneath the animal.
RightTemp Troubleshooting
Problem
Possible Cause
Solutions
Pad is not heating.
Incorrect settings
• Ensure that all settings are properly enabled
for warming and target temperatures are
selected.
Incorrect or loose
connections
• Ensure that the connection at the back of
the PhysioSuite is tight.
Pad is overheating.
Incorrect sensor placement
• Place pad sensors directly beneath the
animal.
• Place animal sensor rectally only.
Incorrect settings
• Ensure that all settings are properly enabled
for warming and target temperatures are
correctly selected.
• Enable a “Maximum Pad Temp” when type
of warming is “Body temp regulated” to
prevent overheating.
“Lost Sensor” alarm
The sensor has been
disconnected.
• Reconnect the temperature sensors.
Sensor failure
• Contact Kent Scientific for assistance.