Procedure For Use
1. Touch Vent to open the RoVent Run Screen.
2. Touch Start Vent to begin ventilation.
Caution:
If you choose to alter the tubing length, be sure to
recalibrate the deadspace. See page 7.
Caution:
Do not start the ventilator before the animal is
intubated and connected to the ventilator tubing.
3. Enter the animal weight.
The RoVent calculates the respiration rate and tidal volume
using the animal’s body weight.
To change these settings, turn Dial A or Dial B. Touch Setup
from the Vent Run Screen to make other changes to Respiratory
Rate, Target Pressure, Tidal Volume, or other advanced settings.
4. From the Vent Run Screen, touch Stop Vent to stop air flow.
Trigger Command Settings
You can assign RoVent-specific functions to either a Dial press or an external signal via
remote cable.
To set a new Trigger Command:
1. From the Vent Run Screen, touch Setup, then More Ventilator.
2. Touch Triggers Menu.
3. Turn Dial B to choose an action.
Some actions will require additional settings. Please contact Kent Scientific if
you need assistance in enabling a trigger function.