4. In setup mode, press to select default option
1. Press to confirm calibration or chosen option
2. Toggles between conductivity, TDS and salinity measurement modes.
DISPLAY
The KS-SAL-20 is built with a clear LCD display that shows the measured value, various mode indicators and units.
Below is a summary to describe the meaning behind each notification on the display.
Indicator
Description
MEAS
Indicates tester is in the measurement mode
CAL
Indicates tester is in calibration mode
SETUP
Indicates tester is in SETUP mode
HOLD
Indicates the HOLD function is active and the current displayed value is on hold.
ATC
Indicates temperature compensation is enabled
REFILLING ELECTROLYTE SOLUTION FOR DO PROBE
1. Remove the packaging of the DO probe. Unscrew its cap from the probe as shown from the image below:
2. Fill the membrane cap halfway with the electrolyte solution
3. Screw back the membrane cap on the probe.
4. Make sure the cathode of the probe makes contact with the membrane cap. Make sure there are no bubbles found
on the electrolyte solution.