GENIUS

9

Infrared Tympanic Electronic Thermometer

Site text

Pressing the °C/°F button when in this mode turns the body site text labels on or off.   

 

 

 

 

The labels will remain on when an “X” appears inside the box icon, and the text will remain off when the box is empty.

���

���

���

����

��

Section VI — Preventative Maintenance 

Functional checks must be performed by technicians trained by Tyco Healthcare/Kendall and based on the procedure in the GENIUS 

2 Infrared Tympanic Electronic Thermometer Checker Operation Manual.  Contact your Tyco Healthcare/Kendall representative for 

details.  Tyco Healthcare/Kendall recommends functional checks at 12 month intervals.  Harsh use or environmental conditions may 

require reduced functional check intervals at some facilities.  If the unit is dropped or abused, check the unit prior to next use.  

Section VII —  Cleaning 

•  The GENIUS 2 Thermometer body may be wiped clean with a damp cloth.  Water temperature should not exceed 55°C (130°F).   

Do not soak, rinse, or submerge the GENIUS 2 Thermometer under water.  

•  The probe tip can be cleaned with a lint free swab.  If the probe tip becomes soiled, it can be cleaned with a dampened swab.  After 

the foreign matter has been removed, the window at the end of the probe tip should be cleaned with a lint free swab or lens wipe.  

The thermometer lens must be free from fingerprints and/or smudges for proper operation.

•  A mild detergent may be added to the water.  Do not use cleaners such as Spray-Nine™*, Phisohex™*, Hibiclens™*, or 

Vesta-Syde™* disinfectants, as they may result in damage to the thermometer case.

•  Occasional use of a 10:1 water and hypochlorite mixture or a damp isopropyl alcohol wipe or Cidex™* or ManuKlenz™* or VIROX™* 

or CaviWipes™* cleansing agents is acceptable.  However, prolonged or repeated use of these chemicals may result in damage to 

the thermometer case and display area.

•  Use of a cloth or sponge is recommended for cleaning.  Never use an abrasive pad or an abrasive cleaner on the GENIUS 2 Thermometer.

ENIUS 2 Thermometer

ENIUS 2 Thermometer

• The thermometer is non-sterile.  Do not use ethylene oxide gas, heat, autoclave, or any other harsh method to sterilize this thermometer.
•  Allow sufficient time for drying before attempting to use the thermometer.

Section VIII  — Battery Replacement

The GENIUS 2 Infrared Tympanic Electronic Thermometer batteries should be replaced when the low battery icon is shown on the LCD 

display.  After the low battery icon is displayed, the user will be able to take approximately 100 temperatures before the dead battery 

icon is displayed and the unit will not allow a temperature to be taken.  To replace the batteries, access the battery compartment by 

unscrewing the battery door that retains the battery cover.  Note the polarity of the installed batteries. Remove the old batteries and 

install fresh batteries, ensuring the correct polarity.  Reinstall the battery cover and secure it with the screw.

Section IX — Mounting Instructions

If using the wall or cart mount options (sold separately) for the base unit, consult the installation sheet provided with the respective system.

Summary of Contents for GENIUS 2

Page 1: ...por Infra Vermelhos Manual de funcionamento e assistência Tympaaninen sähkökäyttöinen infrapunalämpömittari Käyttö ja huolto opas Infrarødt elektronisk tympanisk termometer Betjenings og servicehåndbog Ηλεκτρονικό θερμόμετρο υπερύθρων τυμπάνου Εγχειρίδιο λειτουργίας και συντήρησης Infračervený bubínkový elektronický teploměr Uživatelská a servisní příručka Infravörös timpanikus elektronikus hõmérõ...

Page 2: ...y Toggle 5 Sleep Mode 5 PulseTimer Mode 6 Thermometer Display Icons and Alarms 6 Biotech Mode 7 Section VI Preventative Maintenance 9 Section VII Cleaning 9 Section VIII Battery Replacement 9 Section IX Mounting Instructions 9 Section X Troubleshooting 10 Section XI Specifications 10 Section XII Customer Service 12 Section XIII Warranty 13 This product contains software soley owned byTyco Tyco gra...

Page 3: ...uisition that meets CEN and ASTM standards see Section XI Peak Select System Temperature measurement range of 33 0 C to 42 0 C 91 4 F to 107 6 F After a temperature has been acquired the C F button alternates the displayed temperature between C and F Audible and visual indication of completed temperature acquisition Disposable single use probe covers aid in the prevention of a cross contamination ...

Page 4: ...peraturebeforeuse Excessive eardrum scarring may cause lowered temperature readings Under normal conditions ear wax does not affect accuracy However cerumen plugs or impactions containing debris can lower the temperature measurement by several tenths of a degree Always wait at least two minutes before taking another measurement in the same ear Used probe covers must be treated as infectious biolog...

Page 5: ...perature measurements with GENIUS 2Thermometers The factory calibration and final determination of an acceptable GENIUS 2Thermometer is dependent on the use of acceptableTyco Healthcare Kendall GENIUS 2Thermometer ACCUSYSTEM final determination of an acceptable final determination of an acceptable probe covers The use of generic probe covers or other probe covers not produced byTyco Healthcare Ken...

Page 6: ...US 2Thermometer The use of probe covers produced by any company except Tyco Healthcare Kendall will result in erroneous temperature measurement In order to aid in infection prevention and control always install a new probe cover prior to taking a temperature The probe covers are contained in a cassette that is located in the thermometer base To load a probe cover on the thermometer firmly insert t...

Page 7: ...rmometer to base for storage Temperature Recall After a temperature has been acquired the thermometer will enter off mode after approximately 10 seconds The temperature can be recalled by pressing and releasing the scan button or by pressing and holding the C F button Temperature Display Toggle C or F Whenatemperatureisinthedisplay theusermaypressandholdthe C FbuttontotogglebetweendegreesCelsiusan...

Page 8: ...ion will return the thermometer to off mode 4 At the end of the 60 seconds the thermometer will wait two seconds and then enter sleep mode 5 Return the thermometer to base for storage Thermometer Display Icons and Alarms Thermometer operation is communicated to the user via the LCD display and an audible device contained inside the thermometer case After the probe cover is installed or the batteri...

Page 9: ...factory calibration parameters are also stored in nonvolatile memory The factory default settings are shown below Temperature mode C unlocked Site mode Ear Site text On The biotech mode is accessed by pressing and holding the timer and C F buttons for four seconds All LCD segments will light for one second the thermometer will issue a single beep and the display will show scrolling dashes Pressing...

Page 10: ...GENIUS 2 8 InfraredTympanic ElectronicThermometer Temperature mode Site mode ...

Page 11: ...ater Do not use cleaners such as Spray Nine Phisohex Hibiclens or Vesta Syde disinfectants as they may result in damage to the thermometer case Occasional use of a 10 1 water and hypochlorite mixture or a damp isopropyl alcohol wipe or Cidex or ManuKlenz orVIROX or CaviWipes cleansing agents is acceptable However prolonged or repeated use of these chemicals may result in damage to the thermometer ...

Page 12: ...is located in Section XII Customer Service The GENIUS 2Thermometer advisory and alarm conditions are described in theThermometer Display Icons and Alarms subheading of SectionV Instructions for Use Section XI Specifications Clinical accuracy characteristics and procedures are available from the manufacturer on request To verify accuracy use a certified blackbody as specified in EN 12470 5 2003 Ann...

Page 13: ...torage Temperature Range 25 C to 55 C 13 F to 131 F up to 95 RH non condensing If the unit is stored at extremes it is recommended that the unit be checked on the field calibration checker or at the factory before returning to service Clinical Repeatability Meets Section A 5 of EN 12470 5 2003 E perTyco Healthcare Kendall technical report Data is available from Tyco Healthcare Kendall on request R...

Page 14: ...ccount for the known variance in measurements taken due to human factors Section XII Customer Service Tyco Healthcare Kendall will make available on request circuit diagrams component part lists descriptions or other information to assist qualified institutional technical personnel in repairing parts of this device designated byTyco Healthcare Kendall as repairable In the event that it is necessar...

Page 15: ...of of original purchase will be required Tyco Healthcare Kendall does not assume any liability for loss arising from unauthorized repair misuse neglect or accident Removal defacement or alteration of serial lot number voids warranty Tyco Healthcare Kendall disclaims all other warranties expressed or implied including any implied warranty of merchantability or fitness for a particular purpose or ap...

Page 16: ... Handleiding voor bediening en onderhoud Termómetro Auricular Electrónico por Infra Vermelhos Manual de funcionamento e assistência Tympaaninen sähkökäyttöinen infrapunalämpömittari Käyttö ja huolto opas Infrarødt elektronisk tympanisk termometer Betjenings og servicehåndbog Ηλεκτρονικό θερμόμετρο υπερύθρων τυμπάνου Εγχειρίδιο λειτουργίας και συντήρησης Infračervený bubínkový elektronický teploměr...

Reviews: