Keller HPX Operating Instructions Manual Download Page 7

shown, and press ENTER to activate. 

You can now use SELECT to choose 

the function you want, and ENTER to 

execute the function. Depending on 

the  current  setting,  the  first  function 

to be shown is either 

min/max disp

 or 

LEAK disp

.

Leak measurement function  
Leak mode is used to determine 

the  pressure  change  over  a  defined  

period, which can be adjusted. The 

unit to be tested must be connected 

to the high pressure calibrator on the 

pressure side.

Start leak measurement
Activate the

 

MANO

 menu. The display 

shows 

Leak  dISP

. Press the ENTER 

button and then the SELECT button.

 

Press ENTER to confirm

 

Leak Start

The leak measurement starts, and the 

display alternates between the current 

leak time and the pressure change 

measured thus far.

Active leak measurement
During leak measurement, the lower 

part of the display alternates each sec-

ond between the measurement time 

that has now elapsed [mm:ss] and the 

pressure change measured thus far.

End leak measurement early
To end a leak measurement early, 

press the ENTER button and confirm 

the ”

Leak  Stop

” display by pressing 

ENTER.

Leak measurement completed
If the leak measurement time has 

elapsed or if the measurement was 

manually ended ahead of time, 

the display alternates between the 

elapsed leak measurement time and 

the measured pressure change. 

Set leak measurement time
The leak measurement time is preset 

to 10 minutes in the factory, and it can 

only be changed with the ”Mano Con-

fig” software.

(

 Software for calibrators)  

MANO / "Continuous" function 
Automatic switch-off function (the de-

vice switches off automatically 15 mi-

nutes after a button was last pressed). 

Leak measurements are canceled by 

the automatic switch-off function if the 

measurement time is more than the 

switch-off time.

Cont  on

: Disables the automatic 

switch-off function

Cont  oFF

: Enables the automatic 

switch-off function

If the "Continuous" function is en-

abled, 

Cont

 flashes on the display.

 

Mano mode

 

Launch Leak 

mode by press-

ing the ENTER 

button  

7.  Maintenance / disposal

Venting the pressure system
Release the pressure completely and 

then open the drain valve (10) and the 

screwed sealing plug (7). Screw the 

screw compressor (11) in completely. 

Pump steadily with the manual boos-

ter pump (8) to clear the system of air. 

When no more bubbles come out of 

the recirculation pipe (6), close the 

drain valve (10). 

Changing the oil
We recommend that you have 

KELLER change the oil. The entire 

system is cleaned at the same time.

Only use type HLP 22 BP hydraulic oil.

Battery 
The pressure calibrator is powered 

by a 3 V button-cell battery (behind 

the display). If the battery is low, the 

battery symbol on the display 

  

lights up.

Replacing the battery
Please switch the device off. Turn the 

display section ring beyond the limit 

stop until it is released from the hous-

ing section (turn through about 180°).  

Open the battery compartment and 

change the battery (type CR 2430).

 

Start the leak 

measurement 

with ENTER

ENTER

ENTER

ENTER

KELLER

SE

L E C T

E N

T

E

R

www.

kelle

r-d

ruc

k.c

h

swis

s m

ade

Max./M

in.

17

  Unit of pressure

 Current  

measured value

Leak run-

time in 

mm:ss

  Unit of pressure

 Current  

measured value

Leak run-

time in 

mm:ss

Pressure change

within leak  

runtime

Pressure change

within leak  

runtime

(8)

(10)

(7)

(11)

Summary of Contents for HPX

Page 1: ...Operating instructions for the high pressure calibrator HPX...

Page 2: ...3 Operating the HPX calibrator 15 4 Description of the functions 15 5 Menu navigation for calibrators 16 6 Commissioning 16 7 Maintenance disposal 17 8 Software for calibrators 18 Technical data 19 Sp...

Page 3: ...r K 114A PC connection RS485 2 Device base 3 Display 4 SELECT and ENTER buttons 5 Oil chamber 6 Recirculation pipe 7 Screwed sealing plug 8 Manual booster pump 9 Pressure connection for test object wi...

Page 4: ...A pressure resistant connection for the test object is required in order to use the high pressure calibrator The pressure connection for the test object is already screwed to the pressure distributor...

Page 5: ...y 2 Open the drain valve 10 IMPORTANT Do not open if there is high pressure in the system If you can no longer reach the desired pressure please consult the section on Maintenance to find out how to v...

Page 6: ...iven below in addition to the diagram above Selecting functions The individual sub functions are called up from the MANO menu Keep the SELECT button pressed until MANO is SELECT button The SELECT butt...

Page 7: ...switch off function the de vice switches off automatically 15 mi nutes after a button was last pressed Leak measurements are canceled by the automatic switch off function if the measurement time is m...

Page 8: ...very K 114A or can be downloaded free of charge at www keller druck com Settings on the high pressure calibrator with the ManoConfig software Device settings such as the leak measurement time or the s...

Page 9: ...CR2430 Battery lifetime 2000 h in continuous operation Hydraulic oil HLP 22 BP Dimensions L x W x H 315 337 x 155 x 148 mm Degree of protection IP 65 Selectable pressure units bar mbar hPa kPa MPa PS...

Page 10: ...Test object adapter G 1 4 M G 3 8 F 506810 0028 x Test object adapter G 1 4 M G 1 2 F 506810 0013 x Sealing ring G 1 8 508635 0001 x x Sealing ring G 1 4 508635 0002 x x CrNi filter 307025 2011 x x Va...

Page 11: ...irements for the electromagneticcompatibility whichare established in the directive of the European Community 2004 108 EC This declaration is valid for products of this Series marked with the CE sign...

Page 12: ...KELLER AG f r Druckmesstechnik St Gallerstrasse 119 CH 8404 Winterthur Tel 41 0 52 235 25 25 Fax 41 0 52 235 25 00 www keller druck com November November Novembre 2012...

Reviews: