Keller HPX Operating Instructions Manual Download Page 4

14

Notes on the operating instructions

-  The operating instructions are in-

tended for specialist workers and 

trained personnel. 

-  Before each stage of work, read the 

relevant notes and warnings care-

fully, and keep to the sequence as 

stated. 

-  Pay particular attention to the sec-

tion on ”General safety warnings”. 

If you have any problems or questions,  

please contact your supplier or 

consult KELLER directly.

1.  Description of the device

General description
The high pressure calibrator enables 

pressure to be generated by means of 

the integrated pressure pump, up to 

700 bar relative.

The measurement technology incor-

porated into this device allows ac-

curate measurement and documen-

tation of the characteristic of a test 

object that is connected to it. The 

measured pressure progression can 

be displayed, evaluated and saved 

with a computer monitoring program 

(CCS30). 

The calibrator is operated with the 

two function buttons SELECT and 

ENTER, located directly below 

the display. The calibrator itself is  

powered by a 3,0 V battery, but power 

can also be supplied externally via the 

K-114A interface converter. Test ob-

jects (transmitters or pressure switch-

es) must be supplied from an external 

source.

Pressure range for the display
High pressure calibrators are them-

selves calibrated with the ambient air 

pressure as the zero point reference. 

The Zero function allows any desired 

pressure value to be set as the new 

zero point reference. 

To take relative measurements, 

the high pressure calibrator is 

 

zeroed at ambient pressure (

Set 

Zero

). To reset the pressure zero 

point to the factory setting, use the  

RES Zero 

function (reset zero).

Commissioning
A pressure-resistant connection for 

the test object is required in order to 

use the high pressure calibrator. The 

pressure connection for the test object 

is already screwed to the pressure  

distributor of the high pressure calibra-

tor so that it is pressure resistant when 

it leaves the factory, and it must not be 

dismantled.

Recommended torque for the test 

 

object pressure connection: 30 Nm

IMPORTANT!

Nothing must adhere to the surface of 

the test object (no oil, grease, water, 

etc). Impurities could pass through the 

adapter to reach the high pressure 

calibrator and damage it.

Overpressure
If the pressure exceeds the measuring 

range by more than 20%, the measur-

ing cell or the mechanism of the high 

pressure calibrator may be destroyed.

Recalibration 
The recalibration cycle depends on 

the conditions of use. Recommended 

recalibration cycle: 1 year. 

Scope of delivery
-  1 calibrator (including oil filling)

-  1 carrying case

-   1 test record (5 points)

-   1 test object adapter G 1/4”–G 1/8”

-   1 test object adapter G 1/4”–G 1/2”

-   1 Allen key 

-   1 spare battery, type CR2430 (3,0 V)

-  1 set of operating instructions

-  1 USB interface converter, K-114A  

-  1 KELLER software CD

Intended use
The high pressure calibrator (HPX) 

may only be used to generate pres-

sure with the type HLP 22 BP hydrau-

lic oil that is supplied with the product. 

Use of the calibrator with other media 

will damage it. The operational safety 

of the device supplied is guaranteed 

only if it is used as intended. The limit 

values as stated (see page 19: ”Tech-

nical data”) must never be exceeded. 

Before installing the high pressure 

calibrator, check that it is suitable for 

your applications.

2.  General safety warnings

The current national regulations on 

accident prevention and workplace 

Operating instructions for the

high-pressure calibrator (HPX)

October 2012

Scope of delivery

High pressure calibrator

Summary of Contents for HPX

Page 1: ...Operating instructions for the high pressure calibrator HPX...

Page 2: ...3 Operating the HPX calibrator 15 4 Description of the functions 15 5 Menu navigation for calibrators 16 6 Commissioning 16 7 Maintenance disposal 17 8 Software for calibrators 18 Technical data 19 Sp...

Page 3: ...r K 114A PC connection RS485 2 Device base 3 Display 4 SELECT and ENTER buttons 5 Oil chamber 6 Recirculation pipe 7 Screwed sealing plug 8 Manual booster pump 9 Pressure connection for test object wi...

Page 4: ...A pressure resistant connection for the test object is required in order to use the high pressure calibrator The pressure connection for the test object is already screwed to the pressure distributor...

Page 5: ...y 2 Open the drain valve 10 IMPORTANT Do not open if there is high pressure in the system If you can no longer reach the desired pressure please consult the section on Maintenance to find out how to v...

Page 6: ...iven below in addition to the diagram above Selecting functions The individual sub functions are called up from the MANO menu Keep the SELECT button pressed until MANO is SELECT button The SELECT butt...

Page 7: ...switch off function the de vice switches off automatically 15 mi nutes after a button was last pressed Leak measurements are canceled by the automatic switch off function if the measurement time is m...

Page 8: ...very K 114A or can be downloaded free of charge at www keller druck com Settings on the high pressure calibrator with the ManoConfig software Device settings such as the leak measurement time or the s...

Page 9: ...CR2430 Battery lifetime 2000 h in continuous operation Hydraulic oil HLP 22 BP Dimensions L x W x H 315 337 x 155 x 148 mm Degree of protection IP 65 Selectable pressure units bar mbar hPa kPa MPa PS...

Page 10: ...Test object adapter G 1 4 M G 3 8 F 506810 0028 x Test object adapter G 1 4 M G 1 2 F 506810 0013 x Sealing ring G 1 8 508635 0001 x x Sealing ring G 1 4 508635 0002 x x CrNi filter 307025 2011 x x Va...

Page 11: ...irements for the electromagneticcompatibility whichare established in the directive of the European Community 2004 108 EC This declaration is valid for products of this Series marked with the CE sign...

Page 12: ...KELLER AG f r Druckmesstechnik St Gallerstrasse 119 CH 8404 Winterthur Tel 41 0 52 235 25 25 Fax 41 0 52 235 25 00 www keller druck com November November Novembre 2012...