Absolute measurement method (SG)

With the absolute measurement method, the prevailing total pressure on the level sen-

sor  is  determined  (=  air  pressure  +  hydrostatic  water  pressure). To  compensate  for 

the influence of fluctuations in air pressure, a second data logger is positioned on the 

surface as a barometer.

Advantage: 

very robust, impervious to ambient humidity

Disadvantage:  no integrated air pressure compensation 

Absolute/absolute measurement method (A/A)

Total pressure and air pressure are each determined by means of a pressure sensor 

(A/A technology) and are calculated mathematically in the device by subtracting the two 

measured values. A sensor cable without a capillary tube is used for this measurement 

method, and this is why AA systems are highly resistant to the ambient humidity prevail

-

ing at the measuring point.

Advantage: 

very robust, impervious to humidity or flooding

Disadvantage:  slight reduction in measurement accuracy because of measurements  

 

by two sensors 

Level measurements in closed receptacles

For  level  measurement  in  closed  receptacles/vessels, 

the internal pressure in the vessel is measured instead of 

the air pressure. The A/A measurement method is usually 

selected on account of the humidity prevailing inside the 

vessel. 

Air pressure

Internal

pressure

Water 

pressure

Summary of Contents for DCX Series

Page 1: ...1 keller druck com November 2013 OPERATING INSTRUCTIONS OPERATING INSTRUCTIONS DCX DATA LOGGER DCX DATA LOGGER...

Page 2: ...and replacement Disposal of old electrical equipment and batteries rechargeable batteries Fitting instructions 11 Requirements for the location Installation in level tubes Water level configuration I...

Page 3: ...mH2O 500 meters DCX 16 SG absolute 0 8 11 bar abs 0 100 mH2O 500 meters DCX 18 absolute 0 8 11 bar abs 0 100 mH2O DCX 18 VG relative 0 10 bar rel 0 100 mH2O 500 meters DCX 18 SG absolute 0 8 11 bar a...

Page 4: ...eable battery completely welded in screw on cable Series DCX 22 probe diameter 22 mm AA version high accuracy version customized versions on request Series DCX 25 PVDF probe diameter 25 mm for use in...

Page 5: ...ed capillary tube Surface of water AA version VG version SG version Level sensor P1 medium pressure bar TOB1 medium temperature C Air pressure sensor P2 air pressure bar TOB2 air temperature C Referen...

Page 6: ...ways measured The pressure difference determined in this way corresponds to the hydrostatic water pressure As a result of this measurement method the level measurement becomes independent of weather i...

Page 7: ...A technology and are calculated mathematically in the device by subtracting the two measured values A sensor cable without a capillary tube is used for this measurement method and this is why AA syst...

Page 8: ...intervals Data memory size 2 Mbit Type Interval Number of channels Measurement channel Recording period DCX 22 1 sec 1 Total pressure 15 h 54 min DCX 22 1 min 1 Total pressure 21 h 7 h DCX 22 1 h 1 T...

Page 9: ...uct datasheet Do not drop or throw the device Do not attempt to modify the device The sensor pressure diaphragm is sensitive to contact Do not press the diaphragm in by hand or damage it with sharp ob...

Page 10: ...e and shall not pay com pensatory damages in case of data loss Data stored in the product may be lost if the product is repaired or replaced You should always create a backup of all data stored in the...

Page 11: ...a dry area or use moisture absorber cartridges Installation in level tubes Place the cap lock in position and screw the two grub screws in to prevent it from turning anti theft protection requires a...

Page 12: ...vel surface of water e g using a light plummet and is entered in the water level configuration in the Logger 5 software together with the density of the medium The system length is the sum of the hydr...

Page 13: ...l measure ment will be affected Standard version The pressure compensation aperture is located be hind the removable protective tube which is perforated This opening for the cable capillary is protect...

Page 14: ...PC time To do this enable the Synchronize device time with PC time field in the Logger 5 software and then register the configuration on the DCX The DCX s internal clock does not take account of summe...

Page 15: ...ly and the memory can not be managed or deleted by the user Continuous circular buffer The Continuous function circular buffer releases the data logger s entire storage space for the measurement in pr...

Page 16: ...dis charged battery is inserted To ensure that the logger operates reliably it is advisable to replace the battery as soon as its capacity falls below 30 or after 5 years of operation Changing the ba...

Page 17: ...ous long lasting immersion Please consult the datasheet for the precise specifications Check that all sealing rings are clean and undamaged and that the battery compart ment cover is firmly closed so...

Page 18: ...is in proper working order Moisture absorber cartridges optional The condition of the moisture absorber cartridge can be determined from the coloration of the granulate Blue granulate is dry pink gran...

Page 19: ...e sensor diaphragm is very sensitive Do not touch the diaphragm Cleaning the air pressure sensor AA versions To clean the pressure sensor as thoroughly as possible remove the knurled nut battery sleev...

Page 20: ...r calibration The following test options may be chosen simple check on the DCX without documentation internal check with documentation 5 or 11 point test protocol or an internal check followed by ex t...

Page 21: ...esult in misinterpretations of the measurement result on account of the heterogeneous temperature distribution in bodies of still water In normal cases the temperature distribution is heterogeneous so...

Page 22: ...2 13 14 7 8 1 12 14 7 8 No Designation 1 interface plug RS485 2 Sealing ring for interface plug 3 Sealing ring for battery compartment 4 DCX 22 38 VG AA lithium battery 4 DCX 16 AAA lithium battery 5...

Page 23: ...sent on VG versions 14 Screw nut 15 Circlip 16 Sealing ring for battery compartment DCX 25 PVDF 15 5 x 1 5 NBR 17 Sealing ring for battery compartment DCX 25PVDF 16 x 2 Viton 18 Plastic housing compon...

Page 24: ...nen Keep dry before opening Vor ffnen Datenlogger trocknen Keep dry before opening DCX 22 AA DCX 22 VG 1 12 3 5 4 6 7 8 11 15 1 13 14 3 4 3 5 7 8 11 15 12 Vor ffnen Datenlogger trocknen Keep dry befor...

Page 25: ...ep dry before opening DCX 22 SG 1 14 3 4 3 5 7 8 11 15 12 Vor ffnen Datenlogger trocknen Keep dry before opening DCX 25 PVDF 1 5 16 4 16 17 7 8 18 15 11 Vor ffnen Datenlogger trocknen Keep dry before...

Page 26: ...supplied as an option Suitable for DCX 16 SG VG DCX 18 ECO SG VG DCX 22 SG VG DCX 22 AA DCX 38 VG Dimen sions mm x y Drawing no Product no 30 25 33386 pos 1 506810 0006 40 25 33386 pos 2 506810 0018 4...

Page 27: ...5 506815 0042 6 506815 0043 Adapter ring to match Optional DCX cap lock 3 506810 0085 4 506810 0085 5 506810 0087 6 506810 0078 to install the DCX logger in the level tube an adapter ring 2 pot is req...

Page 28: ...ttery Supplied with product 557005 0017 to match for DCX 16 DCX 16 AA SG VG Minamoto brand lithium 3 6 V AAA Battery Supplied with product 557005 0010 to match for DCX 22 DCX 22 AA SG VG Tadiran brand...

Page 29: ...m Plug type power unit 15 V Supplied with product 309010 0086 USA for K 114M DC plug Input 120 VAC Output 15 VDC 360 mA Cable length 2 m Auto adapter cable Optional 309010 0051 Connection to converte...

Page 30: ...ace plug For type DCX 16 11 mm x 1 5 mm Nitrile O ring interface plug Spare part 508610 0055 For type DCX 18 13 5 mm x 1 5 mm Nitrile O ring interface plug Spare part 508610 0051 For type DCX 22 13 mm...

Page 31: ...d end consumer U Out 11 8 VDC I Outmax 40 mA Optical status and configuration display LED Electrical isolation Total length 1 7 m Software CD and USB connection cable included in scope of delivery Int...

Page 32: ...marked with the CE symbol and are a constituent of this declaration The following standards were used to evaluate the products with regard to electro magnetic compatibility EN 61000 6 1 2007 EN 61000...

Reviews: