Keller CellaCast PA8 Series Quick Manual Download Page 8

Quick Start Guide CellaCast PA 8x

EN

9.1  Setting the sampling time at the PA 83

On delivery the pyrometer is parameterized on a discontinuous pouring stream. If the pouring 

times are < 5 seconds the measuring time needs to be adapted.
The time delay (time after detection of the pouring stream up to the start of the measurement) 

and the measurement time must be less than the minimum pouring time.
Two-colour channel [

Qu

] -> ATD function [

Q.Atd

]

Parameter

Function

Default

[

T.DEL

]

Time delay [s] 

1 s

(

T.ACT

)

Sampling time [s] 

Auto

►  Press [Mode] to access the main menu

 

> The display shows [

Ao

► Press the [▼] until [

Qu

] is displayed

►  Press [Mode]

 

> The display shows [

Q.EPS

► Press [▼] until [

Q.ATD

] is displayed

►  Press [Mode]

 

> The display shows [

t.dEl

]

► Press [▼] until [

t.Act

] is displayed

►  Press [Mode]

 

> The current sampling time or [

Auto

] is displayed

►  Press [▲ or ▼] for 2 second

 

> The display flashes three times

► Press [▲ or ▼] to set the desired sampling time
► Press [Mode] to confirm the parameter

Exit operation

► Wait 30 seconds

or 

►  Press [▲ or ▼] to change to the parameter [

End

]. Then Press [Mode] to change to the 

functional menu.

►  In  the  functional  menu  press  [▲  or  ▼]  to  change  to  the  parameter  [

End

], then press 

[Mode].

If the time delay [

T.DEL

] must also be adjusted, proceed as described above. In this case the 

parameter [

T.DEL

] must be selected and changed accordingly.

9.2  Setting the sampling time at the PA 80/81

The pyrometer is parameterized for continuous measurement on liquid metals.

The measuring time is set to 15 s in the delivery state. After the measuring time has expired, the 

measured temperature is shown on the display or the analogue output. Afterwards, the measur-

ing time restarts automatically after the dead time has elapsed.

Summary of Contents for CellaCast PA8 Series

Page 1: ...PYROMETER CellaCast PA 8x Ident no 1117989 05 2022 QUICK GUIDE...

Page 2: ......

Page 3: ...ivated Do not point the laser beam at any person During pyrometer installation and alignment make sure to avoid the possibility of laser light reflections caused by reflective surfaces All currently v...

Page 4: ...th a maximum pressure of 6 bars are required for that Use oil free and dirt free compressed air If the expected ambient temperatures are 65 C a cooling system is required 1 Target size 2 Pouring strea...

Page 5: ...4 20 mA 0 24 V 0 24 V 0 10 V shielded Digital IN OUT1 IN OUT2 Analogue OUT1 OUT2 Analogue IN 1 2 3 4 5 6 7 8 Pin 1 WH white L Power supply 24V DC Pin 2 BN brown Digital in out 1 Pin 3 GN green Analogu...

Page 6: ...tate instrument alignment to the target spot To activate the laser press the MODE button on the rear panel for 2 seconds Alternatively the laser can be switched on with the interface To focus rotate t...

Page 7: ...nt To do this set the pyrometer to the same distance as the distance between the pyrometer and the pouring stream 9 Parameterization On delivery the pyrometer is parameterized on a discontinuous pouri...

Page 8: ...displayed Press Mode The current sampling time or Auto is displayed Press or for 2 second The display flashes three times Press or to set the desired sampling time Press Mode to confirm the parameter...

Page 9: ...r Press or to change to the parameter End Then Press Mode to change to the functional menu In the functional menu press or to change to the parameter End then press Mode Alternatively the parameters c...

Page 10: ...vity ratio carry out a new control measurement The emissivity ratio is set as follows Press or for 2 seconds The value of the selected emissivity ration is displayed for example 103 0 Press or until t...

Page 11: ...If the period for filing claims has expired you will no longer be able to make any claims for compensation of damage or loss Packing The packages used are made of carefully selected environmentally co...

Page 12: ...or in part for any purpose whatsoever without prior consent of the author This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through a...