background image

K-O2 Series Oxygen Monitor

REV2  12/21

CALIBRATION GASES

Pure nitrogen zero gas and a precise mixture of oxygen and nitrogen (see Table 9) are required to calibrate the 
oxygen sensor.  

An orifi ced calibration adapter is recommended to ensure that the sensor is completely immersed in the calibra-
tion gas without applying higher than ambient pressure to it.  
A complete calibration kit that includes all the required accessories in a convenient carrying case is available from 
your distributor.

Type 

Mixture (by volume) 

Recommended Accuracy 

Comments 

Zero gas 

Pure nitrogen 

 

if supplied from liquid source 
beware of cooling the sensor. 

Span gas 

20.9% oxygen  
balance nitrogen 

± 0.1 % oxygen 

Calibration gas inaccuracy adds 
directly to sensor’s specified 
accuracy error. 

TABLE 9: Required Calibration Gases

Blinking Yellow 

During gas sampling period that starts immediately 

when calibration is initiated. 

Blinking Green 

Successful sampling.  Waiting for user to confirm cal 

gas removal. 

Blinking Red 

Failed calibration attempt.  Waiting for user to 

acknowledge with either a re-try or and exit. 

Green/Yellow 

During ambient equilibration period after successful 

calibration.  

New calibration is applied.

 

Red/Yellow 
 

During ambient equilibration period after failed 
sampling.  

Old calibration is unchanged.

 

TABLE 8: 

 

Meaning of status LED blink patterns during calibration.

 

Kele

Manual

ZERO CALIBRATION PROCEDURE

The Zero calibration procedure below MUST be 
done before the Span calibration.
The progress and status of the calibration process 
is indicated by the color and fl ash-state of the 
front cover status LED (see Table 8).  
Apply the nitrogen (zero) calibration gas to the 
sensor using a calibration adapter, following the 
instructions for the calibration kit being used.  
Ensure that gas is fl owing to the sensor, then press 
and hold the ’ZERO’ button (see Figure 9) for 3 
seconds until the front cover LED starts blinking YELLOW, indicating that gas sampling is in progress.  

Ensure that the calibration adapter remains correctly seated and calibration gas continues to fl ow for the 2 

1. 

minute sampling period.  
At the end of the sampling period, the sensor’s status LED blinks GREEN if the sampling was successful or RED 

2. 

if not. 
If successful (blinking GREEN):

3. 

The gas sampling completed successfully. Turn off  the calibration gas fl ow, remove the calibration 

• 

adapter then press and hold the ‘ZERO’ calibration button until the LED blinks GREEN/YELLOW indicating 
that the calibration gas has been removed, the calibration has been applied and the unit is in standby 
for two minutes while the sensor re-equilibrates to the ambient atmosphere before normal operation 
resumes.  The calibration is complete when the status LED returns to steady GREEN.

OR

If NOT successful (blinking RED): 

4. 

The most likely cause of zero calibration sampling failure is insuffi

  cient gas fl ow or leaks around the cali-

• 

bration adapter failing to completely immerse the sensor in nitrogen.  Verify that calibration gas is still 
fl owing at the required rate (typically about 0.2 liters per minute) and the calibration adapter is properly 
positioned.  
The calibration sampling can be re-started while the LED is blinking RED by again pressing and holding 

• 

the ‘ZERO’ button until the LED blinks YELLOW, then return to step 1 above.
To exit the zero calibration routine preserving the original calibration:  turn off  the calibration gas fl ow 

• 

and remove the calibration adapter, then press and quickly release the ‘ZERO’ button.  The status LED 
will blink RED/YELLOW indicating that the calibration gas has been removed, the original calibration has 
been kept and the unit is in standby for two minutes while the sensor re-equilibrates to the ambient 
atmosphere before normal operation resumes.  The original calibration is completely restored when the 
status LED returns to steady GREEN. 

Mechanical 

Chassis Construction 

Industrial strength, 18 Ga. Gray powder-coated steel.  Pad-lockable 
hinged or screw-on cover style available. 

Weight 

2.0 lbs 

Operating Temperature 

-20 to 50°C 

Operating Humidity 

15 – 90 %RH, non-condensing 

Storage Temperature 

-20 to 20°C    (to minimize sensor degrada

on) 

Case Dimensions (H x W x D) 

Lockable hinge cover: 6.4” x 5.9” x 2.4”   (163.5 x 150.8 x 60.7 mm) 
Screw cover: 6.3” x 5.8” x 2.1” (160.0 x 147.3 x 52.0 mm) 

Sensor Vents 

Natural ven

•l

a

on through 18,  0.1” (2.54 mm) diameter vents 

External Indicators 

Tri-color LED indicates opera

onal status of sensor. 

Knockouts 

4 trade ½” knockouts (1 per side) 

Oxygen Sensor (O2) 

Sensor Type 

Galvanic cell 

Measurement Range 

0 – 25%   

Analog Output Range 

4-20mA (corresponds to 0 to 25%) 

Accuracy 

±0.2%   (Typical) 

Calibration Interval 

6 Months 

Sensor Life 

5 or 10 Years (Typical)  

Recommended calibrated Field-
Replaceable Sensor 

KMOD-O2-25 (5 years) or KMOD-O2-50 (10 years) 

Calibration Kit 

UCK-3 kit 

Electrical 

Operating Power Voltage 

14 – 30 VAC (RMS) or DC 
Isolated power supply; separate transformer not required. 

Power Consumption 

< 5W 

Control Relays  

2 separate SPDT line-voltage-capable relays for 
warning/ven

•l

a

on and alarm outputs. 

UL-rated: 10 Amps max at 120/277 VAC (RMS) or 30 VDC. 

Concentration Reporting Output 

Isolated, powered 4 – 20 mA current loop output. 
4 mA output => 0 % concentra

on.  20 mA => 25% 

Maximum loop resistance:  51

0Ω

 

Termination 

Pluggable screw-terminals for use with 12 AWG or thinner wire 

SPAN CALIBRATION PROCEDURE

The Zero calibration procedure in MUST be done before the Span calibration.
The progress and status of the calibration process is indicated by the color and fl ash-state of the front cover status LED (see Table 
8).  

Apply the SPAN calibration gas to the sensor using a calibration adapter, following the instructions for the calibration kit be-

1. 

ing used.  Ensure that gas is fl owing to the sensor, then press and hold the ’SPAN’ button (see Figure 9) for 3 seconds until the 
status LED starts blinking YELLOW, indicating that gas sampling is in progress.  
Ensure that the calibration adapter covers the sensor completely for the 2 minute sampling period.  At the end of the sam-

2. 

pling period, the sensor’s status LED blinks GREEN if the sampling was successful or RED if not. 
If successful (blinking GREEN): The sampling completed successfully. Turn off  the calibration gas fl ow, remove the calibration 

3. 

adapter then press and hold the ‘SPAN’ calibration button until the LED blinks GREEN/YELLOW indicating that the calibration 
gas has been removed, the new calibration has been applied and the unit is in standby for two minutes while the sensor re-
equilibrates to the ambient atmosphere before normal operation resumes.  The calibration is complete when the status LED 
returns to steady GREEN.

OR

If NOT successful (blinking RED): 

4. 

The most likely causes of span gas sampling failure are:
Insuffi

  cient gas fl ow or leaks around the calibration adapter not completely immersing the sensor in the calibration gas.  

• 

Verify that calibration gas cylinder has not run-out and the calibration adapter is properly positioned.  
The oxygen concentration in the cali-

• 

bration gas is NOT between 20.8 and 
21.0 percent (by volume).

The calibration sampling can be re-started 
while the LED is blinking RED by again press-
ing and holding the ‘SPAN’ button until the LED 
blinks YELLOW, then go to step 1 above.
To exit the span calibration preserving the 
original calibration, press and quickly release 
the ‘SPAN’ calibration button.  The status 

LED will blink RED/YELLOW indicating that the 
calibration gas has been removed, the original 
calibration will be preserved and the unit is in 
standby for two minutes while the sensor re-
equilibrates to the ambient atmosphere before 
normal operation resumes. The calibration is 
complete when the status LED returns to steady 
GREEN.

Reviews: