Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-338
2600BS-901-01 Rev. B / May 2013
Details
This command selects whether the blender waits for any one event (the “OR” mode) or waits for all selected
events (the “AND” mode) before signaling an output event.
Example
trigger.blender[1].orenable = true
trigger.blender[1].stimulus[1] = digio.trigger[3].EVENT_ID
trigger.blender[1].stimulus[2] = digio.trigger[5].EVENT_ID
Generate a trigger blender 1
event when a digital I/O
trigger happens on line 3 or
5.
Also see
trigger.blender[N].reset()
(on page 7-339)
trigger.blender[N].overrun
This attribute indicates whether or not an event was ignored because of the event detector state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Instrument reset
Trigger blender N clear
Trigger blender N reset
Not applicable
Not applicable
Usage
overrun
= trigger.blender[
N
].overrun
overrun
Trigger blender overrun state (
true
or
false
)
N
The blender number (1 to 6)
Details
Indicates if an event was ignored because the event detector was already in the detected state when the event
occurred. This is an indication of the state of the event detector that is built into the event blender itself.
This command does not indicate if an overrun occurred in any other part of the trigger model or in any other
trigger object that is monitoring the event. It also is not an indication of an action overrun.
Example
print(trigger.blender[1].overrun)
If an event was ignored, the output
is
true
.
If an event was not ignored, the
output is
false
.
Also see
trigger.blender[N].reset()
(on page 7-339)