Table 15-1
Recommended verification equipment
Description
Manufacturer/
Model
Accuracy
Digital Multimeter
Agilent 3458A
DC Voltage
1
(2601/2602)
DC Voltage
1
(2611/2612)
DC current
1
90mV:
0.9V:
5.4V:
36V:
190mV:
1.8V:
18V:
180V:
90nA:
0.9µA:
9µA:
90µA:
0.9mA:
9mA:
90mA:
0.9A:
±8ppm
±5ppm
±4ppm
±6ppm
±5ppm
±4ppm
±6ppm
±6ppm
±430ppm
±45ppm
±25ppm
±23ppm
±20ppm
±20ppm
±35ppm
±110ppm
0.5
Ω
, 250W, 0.1%
Precision Resistor
Isotek
RUG-Z-R500-0.1-TK3
Resistance
2
0.5
Ω
:
±125ppm
1. 90-day specifications show full-range accuracy of recommended model used for specified
measurement point.
2. Resistor used to test Model 2601/2602 3A range and Model 2611/2612 1.5A range only should be
characterized to uncertainty shown using resistance function of digital multimeter before use.
15-4 Performance Verification
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Verification limits
The verification limits stated in this section have been calculated using only the
Series 2600 one-year accuracy specifications, and they do not include test
equipment uncertainty. If a particular measurement falls outside the allowable
range, recalculate new limits based both on the Series 2600 specifications and
corresponding test equipment specifications.