10-6 Digital I/O Port
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Digital I/O bit weighting
Bit weighting for the digital I/O lines is shown in
.
Table 10-1
Digital I/O bit weighting
Line #
Bit
Decimal
weighting
Hexadecimal
weighting
1
2
3
4
5
6
7
8
9
10
11
12
13
14
B1
B2
B3
B4
B5
B6
B7
B8
B9
B10
B11
B12
B13
B14
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
0x0001
0x0002
0x0004
0x0008
0x0010
0x0020
0x0040
0x0080
0x0100
0x0200
0x0400
0x0800
0x1000
0x2000
Setting digital I/O values
To set digital I/O values:
1. Press the MENU key.
2. Select GENERAL, and then press ENTER or the Rotary Knob.
3. Choose DIGOUT, and then press ENTER or the Rotary Knob.
4. Select DIG_IO_OUTPUT, and then press ENTER of the Rotary Knob.
5. Set the decimal value as required to set digital I/O line(s) within the range of
0 to 16,383 (see
), then press ENTER or the Rotary Knob.
6. Press EXIT as needed to return to the normal display.
Write protecting digital I/O lines
You can also write protect specific digital I/O lines to prevent their values from
being changed as follows:
1. Press the MENU key.
2. Select GENERAL, and then press ENTER or the Rotary Knob.
3. Choose DIGOUT, and then press ENTER or the Rotary Knob.