Series 2200 Programmable DC Power Supplies Reference Manual
Section 2: General operation
2200S-901-01 Rev. C / March 2016
2-17
Initializing to the default setup
Use the default setup to get the power supply into a default initial state:
1. Remove all leads from the output connectors.
2. Turn on the power supply.
3. Press
Shift
and
Menu (1)
to access the configuration menus.
>Default Set
shows on the
display.
4. Press
Enter
to bring up the default settings menu.
No
and
Yes
show on the display.
5. Press the right arrow key or navigation wheel to select
Yes
.
6. Press
Enter
to enable the default settings.
See the following table for the default settings.
Default setting
Output On/Off
OFF
V-Set
1.000 V
I-Set
0.1000 A
Max Volt Set
Off
OVP Set
Off
Out Time Set
Off
Key Lock
Off
Output Recall
Off
Save Last
On
Key Beep
Off
Knob Lock
Off
Port Mode
Trigger
Trig Source
Manual
RI Mode
Off
DFI Source
Off
Communication
RS232
Setting the current limit
You may set the current limit from 0 A to the maximum current value of each model. The maximum
current rating is shown on the instrument name plate.
1. Press
I-set
.
2. Use the arrow keys, numeric keypad, or the navigation wheel to enter the current limit.
3. Press
Enter
to set the current limit.
Setting the voltage limit
You may set the voltage limit from 0 V to the maximum voltage rating shown on the instrument name
plate.
1. Press
V-set
.
2. Use the arrow keys, numeric keypad, or the navigation wheel to enter the voltage limit.
3. Press
Enter
to set the voltage limit.