Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-243
tspnet.reset()
This function disconnects all TSP-Net sessions.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
tspnet.reset()
Details
This command disconnects all remote instruments connected through TSP-Net. For TSP-enabled devices, this
causes any commands or scripts running remotely to be terminated.
Also see
None
tspnet.termination()
This function sets the device line termination sequence.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No
Usage
type
= tspnet.termination(
connectionID
)
type
= tspnet.termination(
connectionID
,
termSequence
)
type
The termination type:
•
tspnet.TERM_LF
•
tspnet.TERM_CR
•
tspnet.TERM_CRLF
•
tspnet.TERM_LFCR
Details
This function sets and gets the termination character sequence that is used to indicate the end of a line for a
TSP-Net connection.
Using the
termSequence
parameter sets the termination sequence. The present termination sequence is
always returned.
For the
termSequence
parameter, use the same values listed in the table above for type. There are four
possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns (CR or 0x13). For
TSP-enabled devices, the default is
tspnet.TERM_LF
. For devices that are not TSP-enabled, the default is
tspnet.TERM_CRLF
.