Trigger events
Event description
Event constant
No trigger event
trigger.EVENT_NONE
Trigger event blender
N
(1 to 2), which combines trigger events
trigger.EVENT_BLENDER
N
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
Example
print(trigger.tsplinkout[3].stimulus)
Outputs the event that will start action on
TSP-Link trigger line 3.
Also see
trigger.tsplinkout[N].assert()
(on page 8-341)
trigger.wait()
This function waits for a trigger event.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
triggered
= trigger.wait(
timeout
)
triggered
A trigger was detected during the timeout period:
true
No triggers were detected during the timeout period:
false
timeout
Maximum amount of time in seconds to wait for the trigger
Details
This function waits up to
timeout
seconds for a trigger on the active command interface. A
command interface trigger occurs when:
•
A GPIB GET command is detected (GPIB only)
•
A VXI-11 device_trigger method is invoked (VXI-11 only)
•
A
*TRG
message is received
If one or more of these trigger events were previously detected, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
Example
triggered = trigger.wait(10)
print(triggered)
Waits up to 10 s for a trigger.
If
false
is returned, no trigger was detected
during the 10 s timeout.
If
true
is returned, a trigger was detected.
Also see
(on page 8-259)