Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-267
Trigger events
Event description
Event constant
No trigger event
trigger.EVENT_NONE
Trigger event blender
N
(1 to 2), which combines trigger
events
trigger.EVENT_BLENDER
N
Trigger timer
N
(1 to 4) expired
trigger.EVENT_TIMER
N
Analog trigger
trigger.EVENT_ANALOGTRIGGER
External in trigger
trigger.EVENT_EXTERNAL
Example
digio.line[2].mode = digio.MODE_TRIGGER_OUT
trigger.digout[2].stimulus = trigger.EVENT_TIMER3
Set the stimulus for output digital
trigger line 2 to be the expiration of
trigger timer 3.
Also see
(on page 8-52)
(on page 8-54)
Digital I/O port configuration
(on page 3-49)
(on page 8-260)
(on page 8-263)
trigger.ext.reset()
This function resets the edge, logic, and stimulus values for the external in/out port to their default values.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.ext.reset()
Details
This function resets the following attributes to their default values:
•
trigger.extin.edge
•
trigger.extout.logic
•
trigger.extout.stimulus
It also clears
trigger.extin.overrun
.