Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 3: Functions and features
DMM7510-901-01 Rev. B / May 2015
3-49
Digital I/O port configuration
The following figure shows the basic configuration of the digital I/O port.
To set a line high (nom5 V), write a
1
to it; to set a line low (nominally 0 V), write a
0
to it. To
allow an external device to control the state of the line, the line must be set to input mode or
open-drain mode. An attached device must be able to sink at least 50
µ
A from each I/O line.
Figure 121: Digital I/O port configuration
For additional details about the digital output, see the Model DMM7510 specifications (available at
the
Keithley Instruments support website
http://www.keithley.com/support
)).