Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-107
[:SENSe[1]]:<function>:TCouple:TYPE
This command indicates the thermocouple type.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
K
Usage
[:SENSe[1]]:<function>:TCouple:TYPE <identifier>
[:SENSe[1]]:<function>:TCouple:TYPE?
<function>
The function to which the setting applies; see
Functions
<identifier>
B
,
E
,
J
,
K
,
N
R
,
S
, or
T
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]
Details
This command is only applicable when the transducer type is set to thermocouple.
Example
TEMP:TRAN TC
TEMP:TC:TYPE K
TEMP:UNIT C
TEMP:TC:RJUN:SIM 30
Set the transducer type to thermocouple.
Set the thermocouple type to K.
Set the units to Celsius.
Set the simulated reference temperature to 30.
Also see
[:SENSe[1]]:<function>:TCouple:RJUNction:SIMulated
(on page 6-106)
[:SENSe[1]]:<function>:TRANsducer
(on page 6-112)
(on page 2-120)