Section 6: SCPI command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
6-14
DMM7510-901-01 Rev. B / May 2015
*RCL
This command returns the instrument to the setup that was saved with the
*SAV
command.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
*RCL <n>
<n>
An integer from 0 to 4 that represents the saved setup
Details
Restores the state of the instrument from a copy of user-saved settings that are stored in the setup
memory. The settings are saved using the
*SAV
command.
If you view the user-saved settings from the front panel of the instrument, these are stored as scripts
named
Setup0<n>
.
Example
*RCL 3
Restores the settings stored in memory location 3.
Also see
(on page 2-150)
(on page 6-14)
*SAV
This command saves the present instrument settings as a user-saved setup.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
*SAV <n>
<n>
An integer from 0 to 4
Details
Save the present instrument settings as a user-saved setup. You can restore the settings with the
*RCL
command.
Any command that is affected by
*RST
can be saved with the
*SAV
command.
You can save up to 5 user-saved setups. Any settings that had been stored previously as
<n>
are
overwritten.
If you view the user-saved setups from the front panel of the instrument, they are stored as scripts
named
Setup0<n>
.