2
Connector parts
The connectors do not come assembled. The individual parts for each connector are bagged separately. The following table
lists the parts for each connector.
Assembly and wiring
WARNING
All wiring must be rated for the maximum voltage in the system. For example, if 1000V
is applied to the front terminals of the Model 2700/2750, the plug-in module wiring must
be rated for 1000V.
Before handling any connector (or cable assembly), make sure all power has been
removed from it.
NOTE
Refer to the appropriate section of the Model 77xx Series Switching Modules Instruction Manual
for switching module terminal identification.
Use Figure 2 as a guide to assemble and wire the connectors. Keep the following points in mind while assembling and wiring
each connector.
• Use the proper sized grommet for the cable to be installed. Make sure to account for supplementary insulation.
• Make all connections as desired using correct wire size (up to 20 AWG).
• For voltages above 42V, add supplementary insulation around the wiring harness where it enters the cable clamp of the
connector backshell (Figure 2).
Figure 2
Assembly and wiring
Table 1
Connector parts
Qty
Description
Keithley P/N
Qty
Description
Keithley P/N
1
50-pin male D-shell with solder cups: 7703-306
1
25-pin male D-shell with solder cups: 7709-308
1
Housing
—
1
Housing
—
1
Backshell (with grommets)
—
1
Backshell (with grommets)
—
50
Contact pins
—
25
Contact pins
—
2
Jackscrews
—
2
Jackscrews
—
Grommet
Wire (Up to 20AWG)
Supplementary
Insulation
Solder Cup Contact
Jackscrew
(Qty 2)