Operation
2-12
Model 196 DMM connections
Connect the Model 196 or other similar DMM to the matrix
card using the general configuration shown in Figure 2-7.
The VOLTS OHMS HI and LO terminals should be connect-
ed to the desired rows using triax/banana cables. For 4-wire
ohms measurements, the OHMS SENSE HI and LO termi-
nals should be connected to two additional rows using the
same type of cables.
NOTE
For low-level voltage measurements, con-
nect the inner shield of the HI cable to
VOLT OHMS LO to minimize noise.
196 DMM
Connect inner shield to LO for
low-level measurements.
(Modify the cable of Figure 2-4.)
HI
LO
1
2
3
4
5
6
7
8
9
10
11
12
A
B
C
D
E
F
G
H
L
K E ITHL E Y
8 x 12L OW
CUR R ENT MATR IX
COL UMNS
R OWS
S IGNAL
GUAR D
200V P K
200V PK
200V P K
WAR NING:
TIGHTE N MOUNTING S CR E WS
TO E NS UR E P R OP E R
CHAS S IS GR OUND
!
237-BAN-3
237-BAN-3
or Special
Triax
Note: See paragraph 2.5.3
and Figur
e
2-4 for
special triax to banana
cable.
7174A
7174A Matrix Card
Figure 2-7
Model 196 DMM connections
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 7174A
Page 69: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 70: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 71: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 72: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 73: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Page 74: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...