1-14
Getting Started
Model 6485 and 6487 User’s Manual
OPER
Toggles the Vsource state (the Vsource state is displayed by an indicator). (Configured
using the CONFIG >> OPER key sequence.)
V-SOURCE
¹
Increments voltage source value
V-SOURCE
ƒ
Decrements voltage source value
3 Operation keys
COMM
Use to control and modify communication properties (GPIB or RS-232).
DISP
Use to turn display on/off.
TRIG
Trigger measurement(s). Takes 6487 out of idle state. Use also to configure trigger
properties.
LIMIT
Use to perform and create limit tests.
DIGITS
Use to set display resolution.
RATE
Use to select measurement rate.
ß
and
©
Use to control cursor position for making selections or editing values.
SAVE
Use to save present setup to a memory location.
SETUP
Use to restore setup to either GPIB or factory defaults, or to a user memory location.
Also use to modify properties of power on defaults to either GPIB or factory defaults,
or to a user memory location.
STORE
Use to start buffer and modify the number of readings to store.
RECALL
Use to display stored readings (including maximum, minimum, peak-to-peak, average,
and standard deviation). The
¹
and
ƒ
RANGE keys scroll through the buffer, and the
ß
or
©
key toggles between reading number, reading, voltage, and timestamp.
AZERO
Use to control auto zero function (on / off).
DAMP
Use to control damping function (on / off).
EXIT
Use to cancel selection and move back to measurement display.
ENTER
Use to accept selection and move to next choice or back to measurement display.
4 Range keys
¹
Use to select the next higher measurement range. Also use to modify the upper auto-
range limit.
ƒ
Use to select the next lower measurement range. Also use to modify the lower auto-
range limit.
AUTO
Enables/disables autorange.
5 Handle
Pull out and rotate to desired position.
6 VOLTAGE SOURCE OPERATE indicator
This indicator will turn on when the voltage source is in operate (voltage source output turned on). It
will also turn on for about 0.7 sec at power-on to indicate that it is functional.
WARNING
Hazardous voltage may be present on the rear panel V-SOURCE
OUTPUT terminals when the operate indicator is on.
Summary of Contents for 6485
Page 40: ...1 28 Getting Started Model 6485 and 6487 User s Manual...
Page 94: ...7 8 Remote Operation and Commands Model 6485 and 6487 User s Manual...
Page 95: ...A Specifications...
Page 98: ...A 4 Specifications Model 6485 and 6487 User s Manual...
Page 105: ...C ExamplePrograms Programming examples lists and defines three types of programming exam ples...
Page 113: ......
Page 114: ......